Research Article

Development and Validation of a Reversed-Phase HPLC Method for the Estimation of Zolpidem in Bulk Drug and Tablets

Table 2

Intra- and interday assay precision data ( ).

Standard concentration
(μg mL−1)
Measured Concentration
IntradayInterday
mg/tabRecovery (%)RSD (%)mg/tabRecovery (%)RSD (%)

10.84.9398.70.24.9599.051.6
21.65.13102.70.35.12102.501.5
32.45.18103.60.35.09101.800.8

Nominal concentration of sample solution was 20 μg mL−1 and tablets were 5 mg.