Research Article
Development and Validation of a Reversed-Phase HPLC Method for the Estimation of Zolpidem in Bulk Drug and Tablets
Table 3
Results of the data analysis for the quantitative determination of zolpidem by the proposed method.
| Statistical parameter | HPLC |
| Concentration range (μg mL−1) | 2.5–30 | Regression equation | | Correlation coefficient () | 0.9996 | Standard error on estimation () | 0.049 | Standard deviation on slope () | 0.00199 | Standard deviation on intercept () | 0.0356 | Limit of detection (LOD) (μg mL−1) | 0.83 | Limit of quantification (LOQ) (μg mL−1) | 2.5 |
|
|