Research Article

Utility of Charge Transfer and Ion-Pair Complexation for Spectrophotometric Determination of Eletriptan Hydrobromide in Pure and Dosage Forms

Table 3

Evaluation of intraday and interday precision and accuracy obtained by the proposed methods.

MethodsAdded
(μg mL−1)
Intra-dayInter-day
Recovery
%
Precision
RSD %a
Accuracy
RE %
Confidence
limitb
Recovery
%
Precision
RSD %a
Accuracy
RE %
Confidence Limitb

Quinz4.099.500.48 100.250.700.25
8.099.001.09 99.151.30
1299.601.32 99.301.70
1699.800.88 99.400.97
ARS2.099.500.56 99.700.69
4.099.251.15 99.900.75
6.099.801.50 99.300.83
8.099.400.84 7.952 ± 0.0799.401.10
Mo(V)-thiocyanate5.099.400.58 99.800.75
1099.101.03 99.501.46
2099.701.52 99.201.15
3099.500.98 99.600.80

Mean of six determinations. RSD %: percentage relative standard deviation; RE %: percentage relative error.
Mean standard error.