Determination of Retinol, -Tocopherol, Lycopene, and -Carotene in Human Plasma Using HPLC with UV-Vis Detection: Application to a Clinical Study
Figure 3
Recovery experiment: retinol (a), -tocopherol (b), lycopene (c), and β-carotene (d). Values of triplicate assays are recorded. Slopes correspond to the mean recovery 97.0% (S.D. = 2.4%) for retinol, 94.6% (S.D. = 1.7%) for -tocopherol, 91.9% (S.D. 3.6%) for lycopene, and 93.9% (S.D. = 4.2%) for β-carotene.