Research Article

Host-Tailored Sensors for Leucomalachite Green Potentiometric Measurements

Table 2

Main analytical features of the LMG potentiometric sensors in 5.0 × 10−2 mol L−1 phosphate buffer, pH 4.5.

ISESlope (mV decade−1) ( )LOD (mol L−1)LLLR (mol L−1) (mV)C (%)

I 0.994 1.4211
II 0.996 3.543.2
III 0.933 6.085.3
IV 0.997 2.331.6
V 0.998 1.770.95
VI 0.993 8.166.0
VII 0.992
VIII 0.993 5.236.0
VIII 0.972 5.231.8

C : coefficient variation if weekly calibration for  M; : correlation coefficient; LOD: limit of detection; LLLR: lower limit of linear range; : standard deviation of the analytical measure for a concentration in the middle of the calibration curve, considering 3 consecutive calibrations.