Intraday and interday precision and accuracy of GFX (QC) samples.

Day of analysisGFX measured concentration in human urine
LQC (15 ng mL−1)MQC (150 ng mL−1)HQC (400 ng mL−1)

Day 114.73154.57400.96
14.89155.05404.32
15.16154.58399.73
14.99153.24402.41
14.89153.60402.91
14.71154.78403.24

Day 214.68154.43402.69
14.73153.67402.22
14.82153.89401.68
14.55154.26400.98
14.75154.43399.54
14.95154.22398.32

Day 314.89153.62400.25
14.56154.57401.15
14.77154.32402.66
14.63153.51403.56
14.48153.17401.12
14.85153.42401.23

Mean14.78154.07401.61
Standard deviation (SD)0.170.571.55
Precision (CV) (%)1.140.380.39
Accuracy98.53102.72100.40