Research Article
High Throughput Quantitative Bioanalytical LC/MS/MS Determination of Gemifloxacin in Human Urine
Table 2
Intraday and interday precision and accuracy of GFX (QC) samples.
| Day of analysis | GFX measured concentration in human urine | LQC (15 ng mL−1) | MQC (150 ng mL−1) | HQC (400 ng mL−1) |
| Day 1 | 14.73 | 154.57 | 400.96 | 14.89 | 155.05 | 404.32 | 15.16 | 154.58 | 399.73 | 14.99 | 153.24 | 402.41 | 14.89 | 153.60 | 402.91 | 14.71 | 154.78 | 403.24 |
| Day 2 | 14.68 | 154.43 | 402.69 | 14.73 | 153.67 | 402.22 | 14.82 | 153.89 | 401.68 | 14.55 | 154.26 | 400.98 | 14.75 | 154.43 | 399.54 | 14.95 | 154.22 | 398.32 |
| Day 3 | 14.89 | 153.62 | 400.25 | 14.56 | 154.57 | 401.15 | 14.77 | 154.32 | 402.66 | 14.63 | 153.51 | 403.56 | 14.48 | 153.17 | 401.12 | 14.85 | 153.42 | 401.23 |
| Mean | 14.78 | 154.07 | 401.61 | Standard deviation (SD) | 0.17 | 0.57 | 1.55 | Precision (CV) (%) | 1.14 | 0.38 | 0.39 | Accuracy | 98.53 | 102.72 | 100.40 |
|
|