Research Article
A Validated HPLC-DAD Method for Simultaneous Determination of Etodolac and Pantoprazole in Rat Plasma
Table 3
Intraday and interday accuracy and precision of quality control samples of PTZ and ETD obtained by HPLC-DAD.
| Day of analysis | PTZ | ETD | Low QC | Medium QC | High QC | Low QC | Medium QC | High QC | (0.3 gmL−1) | (6 gmL−1) | (12 gmL−1) | (8 gmL−1) | (20 gmL−1) | (40 gmL−1) |
| Day 1 | 0.296 | 5.94 | 13.27 | 8.36 | 21.24 | 42.63 | 0.286 | 5.67 | 12.47 | 8.15 | 20.37 | 41.65 | 0.312 | 6.23 | 12.84 | 8.91 | 20.56 | 42.19 | 0.326 | 6.84 | 12.08 | 7.62 | 19.62 | 39.41 | 0.249 | 6.45 | 11.99 | 7.98 | 19.10 | 43.54 | Day 2 | 0.281 | 5.86 | 12.61 | 9.11 | 20.28 | 39.16 | 0.274 | 5.34 | 11.83 | 7.64 | 21.71 | 39.08 | 0.296 | 6.94 | 11.26 | 8.17 | 19.37 | 40.42 | 0.315 | 6.41 | 12.47 | 8.78 | 19.88 | 40.37 | 0.292 | 6.36 | 13.16 | 8.94 | 19.74 | 41.28 | Day 3 | 0.295 | 6.14 | 12.61 | 9.64 | 21.13 | 38.68 | 0.308 | 5.72 | 11.45 | 9.82 | 20.09 | 39.46 | 0.291 | 6.15 | 11.27 | 8.42 | 20.54 | 39.44 | 0.318 | 5.73 | 11.43 | 8.68 | 21.81 | 42.18 | 0.315 | 6.89 | 12.51 | 8.28 | 19.37 | 41.72 |
| Mean ± SD (gmL−1) | 0.29 ± 0.02 | 6.18 ± 0.48 | 12.22 ± 0.66 | 8.56 ± 0.65 | 20.32 ± 0.85 | 40.75 ± 1.52 |
| Precision (% RSD) | 6.89 | 7.76 | 5.40 | 7.59 | 4.18 | 3.73 |
| Accuracy (% error) | −3.33 | 2.97 | 1.83 | 7.08 | 1.60 | 1.87 |
|
|