Research Article

Trace Elements Analysis in Drinking Water of Meghalaya by Using Graphite Furnace-Atomic Absorption Spectroscopy and in relation to Environmental and Health Issues

Table 1

AAS Vario-6 graphite furnace elements instrumental analytical conditions.

ElementsWavelength (nm)Slit width (nm)Atomization temp. (0°C)Matrix modifiersInterference wavelength (nm)Contamination

Li670.80.82250–2300Possible
Na589.00.51400–1450
Cr357.90.82100–2200NH4H2PO4Fe 358.1, Nb 358.0
Mn279.50.21600–1650Mg(NO3)2 + Pd(NO3)2Mg 279.5, Fe 279.5,
Pb 280.2
Possible
Fe248.30.21850–2050Mg(NO3)2Depressions by Al and Si signal possible
Co240.70.22000–2100Mg(NO3)2
Ni232.00.22100–2300Mg(NO3)2
Cu324.80.81800–1900Ni 324.3, Mn 324.9,
Pd 324.3, Ag 324.8,
Eu 324.8
Possible
Zn213.90.81000–1100Pd(NO3)2Cu 213.9, Te 214.3,
As 214.4, Fe 213.6,
Fe 213.9
Possible
As193.70.82050–2250Ni(NO3)2
Se196.01.21750–1900Ni(NO3)2
Rb780.01.21450–1700
Mo313.30.82550–2600Mg(NO3)2Formation of carbide typical
Ag328.10.81150–1350NH4H2PO4Rh 328.1, Zn 328.2,
Cu 327.4, Y 328.9
Possible
Cd228.80.8900–1200NH4H2PO4 + Mg(NO3)2As 228.9, Fe 228.8
Au242.80.82340–2350Mg(NO3)2Sn 242.1, Co 242.5Depression by Pt and Pd signal are possible
Pb217.00.51200–1350 Pd(NO3)2 + Mg(NO3)2Cu 216.5, Fe 216.7,
Ni 216.6, Sb 217.6,
Pt 216.5
Possible
Ca422.71.22350–2500
Mg285.20.81500–1650
K766.51.21450–1600