Review Article

Chromatographic Methods in the Separation of Long-Chain Mono- and Polyunsaturated Fatty Acids

Table 2

Survey of some selected HPLC conditions useful for separation and identification of MUFA and PUFA from various samples.

HPLC systemMobile phaseColumnDetection modeMatrixRemarksReference

Ag-HPLCIsocratic solvent system: hexane-acetonitrile 99.7 : 0.3 (v/v)Chromspher 5 Lipid column ( mm, Varian, Netherlands)UV-VIS detectorDeodorized fish oil containing EPA, DHA, GLA, LA, ALA, SD, and other PUFADetection levela[63]

HPLC-ECDEthanol-acetonitrile 20 : 80 (v/v)ODS 100RP-18 column,
LiChrospher
250 mm × 4 mm, 5 m phase particle
Electrochemical detector, saturated calomel electrode, detection potential −415 mVHuman plasma, containing various fatty acids including arachidonic acid and linoleic and oleic acidsDetection level: 50–1600 pmoles[80]

HPLC-MS-APCIThe gradient solvent program was used: MeCN/EtCN/DCM
60 : 30 : 10 (v/v/v)  
MeCN/EtCN/DCM 30 : 40 : 30 (v/v/v)
Hichrom column HIRPB-250 AM, 250 mm × 2.1 mm, 5 m phase particlePicolinyl esters were separated by use of quadruple mass spectrometer fitted with an atmospheric pressure chemical ionization sourceDinoflagellate Amphidinium carterae containing VLC-PUFA up nonacosahexaenoic acidDetection levela[81]

RP-HPLC-ELSDThe isocratic elution of MeCN/H2O was used: 97.5 : 2.5 (v/v) flow rate of mobile phase 1.5 ml/minRP column (Alltima, C18, Alltech, 250 mm × 4.6 mm), 5 m phase particleMethyl esters PUFA were detected by evaporative light-scattering detector (ELSD)Marine algae Schizochytrium sp. containing DHA, EPA, and their derivativesDetection level: mg [82]

Ag-HPLCDichloromethane-acetonitrile 100 : 0.025 (v/v)Nucleosil 100-5SA
(250 mm × 4.6 mm); Hichrom, Reading, UK
UV-VIS detector, the working wavelength was 250 nm for the anthrylmethyl and naphthyl derivatives and also 275 nm for the naphthylmethyl derivatives of examined octadecenoic acidReference standards of positional isomers of octadecenoic acid and their derivatives (esters)Detection level: g[83]

Ag-HPLCGradient solvent system consists of hexane-isopropanol-acetonitrile  
350 : 100 : 2.75 (v/v/v)  
350 : 100 : 10 (v/v/v)
Chrompack 5 Lipid column (250 mm × 4.6 mm, Netherlands)UV-VIS detector,
the working wavelength was 206 nm
APCI-MS
Oil produced by a marine microbe containing PUFA (DHA, EPA)Detection levela[84]

aNot described by authors; MeCN-acetonitrile, EtCN-propionitrile, and DCM-dichloromethane.