Research Article

Detection of Pb, Ba, and Sb in Blowfly Larvae of Porcine Tissue Contaminated with Gunshot Residue by ICP OES

Table 1

Main ICP OES parameters.

RF power (W)1350
Coolant gas (L min−1)15
Auxiliary gas (L min−1)0.2
Nebulizer gas (L min−1)1.2
Sample uptake (mL min−1)1.0
ViewingAxial

PbBaSb

Analyte wavelength (nm)220.353233.527206.826
Limit of detection (g L−1)1.490.154.79
Limit of quantification ( L−1)4.970.5015.97