Research Article
Development and Validation of a Simple and Selective Analytical HPLC Method for the Quantification of Oxaliplatin
Table 2
Precision (CV) and accuracy of the HPLC method for OXPt quantification.
| Theoretical concentration (µg/mL) | Experimental concentration (mean ± SD; µg/mL) | CV (%) | Accuracy |
| Intraday () | | | | 0.5 | 0.54 ± 0.01 | 1.75 | 107.33 | 3.0 | 3.00 ± 0.07 | 2.32 | 100.07 | 15.0 | 15.04 ± 0.22 | 1.45 | 100.26 | Interday () | | | | 0.5 | 0.52 ± 0.01 | 1.66 | 104.79 | 3.0 | 2.86 ± 0.12 | 4.17 | 95.28 | 15.0 | 14.82 ± 0.08 | 0.57 | 98.77 |
|
|
CV = (SD/mean); accuracy = (experimental concentration/theoretical concentration) × 100.
|