Research Article

A Simple and Improved HPLC-PDA Method for Simultaneous Estimation of Fexofenadine and Pseudoephedrine in Extended Release Tablets by Response Surface Methodology

Table 8

System suitability study.

ParameterPSUFEXLimit [1416]

% RSD of 0.5920.484% RSD ≤ 2%
% RSD of 0.1920.211% RSD ≤ 2%
Tailing factor ()1.6261.150USP ≤ 2.0
Resolution ()10.94 ≥ 2.0
0.018 μg/mL0.013 μg/mL = 3 : 1
0.055 μg/mL0.041 μg/mL = 10 : 1

. on visual detection.