Limits of quantification (LOQs), linearity range, R2, accuracy, and repeatability parameters evaluated at LOQ.

CompoundLOQ (ng·L−1)Linear range (ng·L−1)R2Accuracy (%)Repeatability CV%

PFBA55–2000.999101.110.7
PFPeA11–2500.99991.710.6
PFBS11–2500.99995.611.5
PFHxA11–2500.99997.48.28
PFHpA11–2500.99997.210.4
PFHxS11–2500.99988.949.7
PFOA11–2500.99999.911.5
PFNA11–2500.99993.013.8
PFOS0.20.2–2500.99985.012.4
PFDA55–2500.99981.012.1
PFUnDA55–2500.999107.211.3
PFDoDA55–2500.999107.49.5