Research Article

Rapid Separation of All Four Tocopherol Homologues in Selected Fruit Seeds via Supercritical Fluid Chromatography Using a Solid-Core C18 Column

Table 2

Linearity, limit of detection (LOD), and limit of quantification (LOQ) of the developed method employing the Kinetex C18 column and the SFC-DAD systema.

Tocopherols
α-Tβ-Tγ-Tδ-T

Standard solutions (R2)0.9985 ± 0.00080.9988 ± 0.00050.9990 ± 0.00060.9991 ± 0.0007
Slope and y intercepty = 41.458x − 1349.4y = 70.363x − 2807.5y = 72.367x − 3144.6y = 70.056x − 2939.1
LOD (ng/μL)29323027
LOQ (ng/μL)88979083

x, concentration (ng); y, peak area (mAU). aTocopherols presented as the absolute amount injected and dissolved in 1 μL of solvent.