Research Article

Simultaneous Determination of Content and Antioxidant Activity of Five Components in Hibiscus mutabilis L by HPLC-QAMS

Table 5

Effects of different instruments and columns on relative correction factors.

InstrumentChromatographic columnRelative correction factor
Protocatechuic acidCatechinRutinLuteolin

Agilent 1260Agilent ZORBAX SB-C180.90450.09170.19961.3902
Agilent TC-C180.92280.09310.20111.4133
Phenomenex C180.91340.09380.20021.4405

Shimadzu LC-2030Agilent ZORBAX SB-C180.89040.08920.19701.4007
Agilent TC-C180.91730.09160.20031.4328
Phenomenex C180.90080.09050.19751.4216

Average0.90820.09170.19931.4165
RSD/%1.311.830.831.35