An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose
Table 2
Difference analysis compared with venous reference under various conditions.
| | | MARD (%) | value | CEG | A | B |
| Total | 469 | 11.67 | ─ | 94.67 | 5.33 | Days of wear | 2 | 80 | 11.53 | 0.25 | 93.75 | 6.25 | 3 | 83 | 9.66 | 95.18 | 4.82 | 6 | 87 | 11.79 | 96.55 | 3.44 | 7 | 88 | 10.89 | 96.59 | 3.41 | 12 | 65 | 13.18 | 95.38 | 4.62 | 13 | 66 | 13.92 | 89.39 | 10.61 | Glucose level | <4 mmol/L | 79 | 16.89 | <0.0001 | 91.14 | 8.86 | 4~10 mmol/L | 292 | 11.58 | 94.52 | 5.48 | >10 mmol/L | 98 | 7.79 | 97.96 | 2.04 | Time interval | 7:00~21:00 | 273 | 10.88 | 0.03 | 95.24 | 4.76 | 21:00~7:00 | 196 | 12.77 | 93.88 | 6.12 |
|
|
Wilcoxon/Kruskal-Wallis test. MARD: mean absolute relative difference; CEG: consensus error grid. |