An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose

Table 3

The CCC index of sensor blood glucose compared with venous blood glucose.

CCC

Overall4690.97 (0.96-0.97)
Glucose level
 <4.0 mmol/L790.49 (0.30-0.64)
 4.0~10 mmol/L2920.87 (0.84-0.90)
 >10 mmol/L980.92 (0.88-0.94)
Days of wear
 2-31630.97 (0.97-0.98)
 6-71750.98 (0.97-0.98)
 12-131310.94 (0.91-0.95)

CCC: concordance correlation coefficient. 95% confidence interval.