Performance of Point-of-Care Testing Compared with the Standard Laboratory Diagnostic Test in the Measurement of HbA1c in Indonesian Diabetic and Nondiabetic Subjects

Table 2

Sensitivity and specificity of the POCT results using the HemoCue HbA1c 501 system based on a 6.5% cut-off point.

HbA1c test by POCT (%)HbA1c test by standard laboratory method (%)
HbA1c≥6.5HbA1c<6.5

HbA1c≥6.54514
HbA1c<6.5148
Total4662

The values in the table represent the number of subjects (). Sensitivity (45 out of 46/97.83%), specificity (48 out of 62/77.42%), positive predictive value (45 out of 59/76.27%), negative predictive value (48 out of 49/97.96%), false positive (14 of 62/22.58%), false negative (1 of 46/2.17%), and accuracy (86.11%).