Research Article

Cross-Camera External Validation for Artificial Intelligence Software in Diagnosis of Diabetic Retinopathy

Table 2

Sensitivity and specificity of VeriSee DR in diagnosing referable DR in three different camera models.

Fundus camera modelSensitivity (95% CI)Specificity (95% CI)

TRC-NW40089.4% (80.6-98.2%)94.2% (92.0-96.4%)
TRC-NW8 series94.7% (89.6-99.8%)90.4% (87.4-93.3%)
Nomyd 8 series95.7% (92.1-99.4%)89.3% (86.0-92.5%)

CI: confidence interval.