Research Article

Development of a Blood Pressure Measurement Instrument with Active Cuff Pressure Control Schemes

Table 3

Measurement results of proposed and commercial BP device of five healthy subjects (unit: mmHg).

SBP/DBP mmHgSubject 1Subject 2Subject 3Subject 4Subject 5Average measurement time
Method

Commercial electronic BP device (1st measurement)115/77132/61113/74103/53107/7838
Commercial electronic BP device (2nd measurement)117/77133/53107/6489/5199/63
Commercial electronic BP device (3rd measurement)116/74137/60100/6287/48101/64
Commercial electronic BP device (8th measurement)121/82123/65105/5987/52105/67
Average116/77131/59106/6491/51103/68
ADM mode (4th measurement)114/75126/70114/6691/64107/6032
ADM mode (5th measurement)119/80125/71109/6784/47110/71
Average116/77125/70111/6687/55108/65
AIM mode (6th measurement)131/66125/65129/8094/52123/6825
AIM mode (7th measurement)121/75139/70114/6586/53112/65
Average126/70132/66121/7290/52117/66