Research Article
Development of an Omnidirectional-Capable Electromagnetic Shock Wave Generator for Lipolysis
Table 1
Average peak pressures of focused shock waves for various specimens.
| Specimen | RO water | 1 cm thick adipose tissue | 2 cm thick adipose tissue | 3 cm thick adipose tissue |
| Operating voltage (kV) | 5 | 5.5 | 6 | 6.5 | 5.5 | 6 | 6.5 | 5.5 | 6 | 6.5 | 5.5 | 6 | 6.5 | Average (MPa) ± SD | 0.87 ± 0.004 | 0.99 ± 0.003 | 1.15 ± 0.067 | 1.38 ± 0.06 | 0.93 ± 2.37 | 1.08 ± 2.57 | 1.29 ± 1.74 | 0.9 ± 1.1 | 1.04 ± 2.77 | 1.25 ± 0.91 | 0.87 ± 1.07 | 1.01 ± 1.19 | 1.21 ± 1.07 | Energy intensity change (%) | n/a | −6.1 | −6.1 | −6.5 | −9.1 | −9.6 | −9.4 | −12.1 | −12.3 | −12.3 |
|
|