Research Article

Development of an Omnidirectional-Capable Electromagnetic Shock Wave Generator for Lipolysis

Table 1

Average peak pressures of focused shock waves for various specimens.

SpecimenRO water1 cm thick adipose tissue2 cm thick adipose tissue3 cm thick adipose tissue

Operating
voltage (kV)
55.566.55.566.55.566.55.566.5
Average
(MPa) ± SD
0.87 ± 
0.004
0.99 ± 
0.003
1.15 ± 
0.067
1.38 ± 
0.06
0.93 ± 
2.37
1.08 ± 
2.57
1.29 ± 
1.74
0.9 ± 
1.1
1.04 ± 
2.77
1.25 ± 
0.91
0.87 ± 
1.07
1.01 ± 
1.19
1.21 ± 
1.07
Energy intensity
change (%)
n/a−6.1−6.1−6.5−9.1−9.6−9.4−12.1−12.3−12.3