Development of an Omnidirectional-Capable Electromagnetic Shock Wave Generator for Lipolysis
Table 2
Average energy intensities of focused shock waves for various specimens.
| Specimen | RO water | 1 cm thick adipose tissue | 2 cm thick adipose tissue | 3 cm thick adipose tissue |
| Operating voltage (kV) | 5 | 5.5 | 6 | 6.5 | 5.5 | 6 | 6.5 | 5.5 | 6 | 6.5 | 5.5 | 6 | 6.5 | Average (mJ/mm2) ± SD | 0.052 ± 0.6e−3 | 0.059 ± 0.4e−3 | 0.068 ± 0.5e−3 | 0.081 ± 0.5e−3 | 0.056 ± 0.28e−3 | 0.064 ± 0.27e−3 | 0.076 ± 0.18e−3 | 0.053 ± 0.19e−3 | 0.061 ± 0.15e−3 | 0.072 ± 0.21e−3 | 0.05 ± 0.29e−3 | 0.057 ± 0.20e−3 | 0.067 ± 0.24e−3 | Energy intensity change (%) | n/a | −5.1 | −5.9 | −6.2 | −10.2 | −10.3 | −11.1 | −15.3 | −16.2 | −17.3 |
|
|