Accurate Rapid Lifetime Determination on Time-Gated FLIM Microscopy with Optical Sectioning

Table 1

Fluorescence lifetimes measured for Coumarin 153 and Erythrosine B solutions in capillary tubes using the RLD technique on epi-fluorescence and HiLo images. The measured values correspond to mean ± standard deviation of the intensity-weighted fluorescence lifetime image. The reference lifetimes were obtained with a time-correlated single-photon counting microscope.

SolutionReference lifetime (ps)Epi-fluorescence image lifetime (ps)HiLo image lifetime (ps)

Coumarin 1533959.73825.6 ± 198.13846.6 ± 463.5
Erythrosine B490.7604.2 ± 40.4613.2 ± 42.0