A Numerical Model of Blood Flow Velocity Measurement Based on Finger Ring

Academic Editor: Antonio Gloria
Received16 Jul 2017
Accepted31 Jul 2018
Published03 Oct 2018

Abstract

Aiming to measure the blood flow velocity in a finger, a novel noninvasive method, i.e., a ring with a heat source chip and a temperature sensor, is designed in this paper. The heat source chip is used to heat the finger and generate heat diffusion between the chip and the temperature sensor. And the temperature sensor is designed to measure the temperature difference. Since the blood flow is the main medium of heat diffusion in bodies, part from the heat energy in the tissue will be taken away by the flowing blood. Therefore, the blood flow velocity can be acquired via its relationship with the temperature difference. Compared to the ultrasound Doppler method and the laser Doppler method, the proposed method guarantees a more convenient operation in more flexible work sites. We also analyze the theory between heat transfer and laminar flow. Finally, several simulations are conducted, and the influence of the relevant factors (i.e., the number of blood vessels, the radius, etc.) corresponding to the simulation results is also discussed.

1. Introduction

Both the hemodynamics and blood rheology are important aspects in the medical research field [1, 2]. The blood flow information measurement is valuable for the cardiovascular diagnosis and some other vascular disease diagnosis. The process in which blood flow reaches the venules through tiny capillaries arterioles is called blood microcirculation. And over 80% substance exchange of a person is completed in microvascular area. The blood microcirculation is present throughout people’s bodies. Therefore, the blood flow information on the microcirculation is particularly significant. Researching blood microcirculation is beneficial for understanding the pathogenesis, understanding the analysis, and determining the pathogenetic condition. Among the blood flow information, the blood flow velocity is one of the main parameters to characterize the microcirculation [36].

There are several traditional blood flow measurement methods, such as the tracer injection method, the plethysmography method, the electromagnetic flow meter based method, etc [711]. Additionally, some of the blood flow sensors and mechanical blood flow meter are also commonly used in biomedical or clinical experiments [1216]. However, the low spatial resolution of traditional methods cannot meet people’s needs any more, in the meantime some methods of them are hard to operate, so the better methods are needed.

Microfluidic technology has emerged as a powerful tool for biological cell manipulation and disease diagnosis [1732]. With the technology diffusion and technology accumulation, a number of the blood flow velocity measurement methods appeared with benefits of high-resolution, nondestructive, and fast test. These methods also become increasingly important in the biomedical field. Among these methods, the Doppler-based method (based on ultrasound and based on laser) is the most commonly used method of the blood flow velocity measurement. The ultrasound Doppler method (the laser Doppler method) uses ultrasonic (laser) to make the blood images within the organization on the screen, and then blood flow velocity is obtained via testing the change of these blood images [3337]. At present, for scientific research and diagnosis, most hospitals are equipped with the Doppler anemometry, which has been improved and applied in clinic for decades. Obviously, the Doppler based methods have been recognized by the broad masses of medical workers. However, the Doppler anemometry is too huge to operate conveniently, and it requires operating personnel of a higher level of knowledge. Besides, the test site is relatively fixed. The Doppler anemometry is not suitable for fingers or some other parts of human bodies, and it is not convenient enough in some work sites (i.e., home, scientific research establishments, etc.).

In this paper, a noninvasive method is proposed to measure blood flow velocity of a finger via heat diffusion between a finger and a ring. The ring consists of a heat source chip and a temperature sensing display device. The temperature difference between these two components indicates velocity information, since the blood flow is the main medium of heat diffusion for a person. Based on actual range of parameters, we assume that the related factors (i.e., the number of blood vessels, radius, etc.) are under the constant conditions, and different blood velocities will necessarily cause diverse temperatures which are obtained from temperature sensor. By using the empirical parameters of the blood, the bone, the adipose, and the finger blood flow rate of normal adults, we also analyze the theory between the heat transfer and the laminar flow, since these two theories are both tightly connected with the overall model [3844]. Finally, we conduct some simulations and give detailed discussions about the simulation results.

2. Method and Theory

The frame diagram of the proposed method is shown in Figure 1. We use multiphysics simulation software—COMSOL Multiphysics to establish the geometric model firstly.

In the human body, three types of blood vessels exist, and any one kind of them has its own function. All these three types of blood vessels can be roughly regarded as tubes with liquid. According to [4547], in addition to the thumb, the digital arteries are on either side of the other four fingers, and they are relatively fixed and straight. The thickness of the proximal artery is relatively symmetrical. In the early stages of modeling, in order to simplify complexity of the calculation, we blur the conception of these three vessels since they have many similar functions and model the numerous capillaries in fingers as a larger vessel. In modeling, due to the smaller dimension of the blood vessels and especially when they are relative to fingers, we set blood vessels as cylinders. Moreover, since the object we study is a section of a finger, the outline of a section of a finger and the bone can both be roughly modeled as cylinders, even though human fingers are not symmetrical. The same way to model the heat source component and the temperature sensing display device, to establish a preliminary model, we simplify them as 1/4 parts from ring cylinders. The overall model diagram is shown in Figure 2.

After the preliminary model is established, we fill the necessary materials into the model. The needed materials are chiefly 5 types: the blood, fat, the bone, metal, and air. According to [3842], the parameters corresponding to these materials are shown in Table 1.


BloodAdiposeBoneAirRing

(J/(kg·K))33000.511001.31.2
(W/(m·K))25030.26911
(kg/m3)21000.32710
(Pa·s)1025.30.0241.29
3854008960

In the method we proposed, we assume that two main components are considered in the ring as the schematic shown in Figure 3. About the heat source component, we assume that the temperature of a heat source is 323.15 K, which is higher than the human finger temperature since the temperature of human fingers is around 303.15 K [48]. Temperature difference leads to temperature transfer in the media. The heat transfers from the heat source component, through finger and finally gets to the temperature sensing component, where the temperature sensor is sensitive to a tiny temperature change. Because a part from the heat will be taken away by the blood flow, the temperature of sensor will be lower than 323.15 K.

In reality, each person has exclusive state of fingers, and people do not need particularly accurate health data in some conditions. Therefore, the relationships between the blood flow velocity and the obtained temperature can be paid attention in those states. These relationships rely on normal ranges of each parameter in truth. The more situations are considered, the better this presupposed ring works.

In heat transfer process, the connected physics fields are the heat transfer field and the laminar flow field. The heat transfer physics field contains two parts that are the heat transfer in solid and the heat transfer in liquid. And the first part is corresponding to the ring, the adipose, and the bone, while the second part is corresponding to the vessels and the blood.

To simulate the process of heat diffusion in the proposed model, the heat transfer interface (interface 1) and the laminar flow interface (interface 2) of simulation software (COMSOL Multiphysics) are mainly used, which are relied on the theory of heat transfer and the theory of laminar flow [49].

Interface 1 contains two parts, the one about fluids is called interface 1.1, and the other one about solids is called interface 1.2 [5052]. The interface 1.2 is used to model the heat transfer process in solids. The areas connected with interface 1.2 are the air, the metal ring, the adipose, the bone, and the blood vessel, which are shown in Figure 4.

The physics involved in interface 1.2 can be described as a temperature equation which is shown below. The interface is used for solving the following equation:

The interface 1.1 is used in the area of the blood vessel. And the areas connected with interface 1.1 are shown in Figure 5. The interface is used for solving the following equation:where two symbols are used to describe the heat flux: (conduction) and (radiation); presents the density; and the thermal conductivity is described as . In equations, the specific heat capacity is presented as , and describes the absolute temperature. is the velocity vector, besides, presents its translational motion. In Equation (1), two parts are on the right side of the equal sign. The first one means additional heat sources, the second one means the thermoelastic damping and accounts for thermoelastic effects in solids:

The operator is the material derivative, as described in the time derivative subsection of material and spatial frames.

In Equation (2), three parts are on the right side of the equal sign. The first one means the heat sources except viscous dissipation.

The second one describes the work produced under changed pressure.where presents the pressure, and the coefficient of thermal expansion is defined as :

The third one represents the viscous dissipation in the fluid:where the viscous stress tensor is presented as . Operation “:” here describes contraction between the tensors, and the details are shown as follows:

Laminar flow is a kind of single-phase fluid-flow, and interface 2 is to describe some parameters under the circumstance. For instance, the velocity and pressure fields in the laminar flow are determined by the scientific regime. A parameter called Reynolds number can be used to describe the state of the flow, and once this parameter is below a certain critical value, the flow remains laminar [5356]. When this parameter is higher, disturbances grow and laminar transits to turbulence [57]. In the model, the interface 2 is used in the area of the blood vessel, which is shown in Figure 5.

The physics behind interface 2 can be described below, and these equations are according to Navier–Stokes equations.where presents the unity tensor, the volume force vector is described as , and the strain-rate tensor is described as :

Conservation of mass is involved in Equation (8), and it is a continuity equation.

Conservation of momentum is involved in Equation (9), and it is a vector equation.

Conservation of energy is involved in Equation (10).

In addition, the other parameters are consistent with parameters in heat transfer interface. Thereinto, the velocity vector is the most important parameter that we pay attention to, which represents blood velocity here. In order to reflect the combined effect in this simulation, we blur the conception of the capillary, the vein, and the artery. For the adult group, we choose an appropriate range of blood velocity, which is from 3 cm/s to 12.6 cm/s. It is based on the empirical values of blood velocity, which ranges from 4.9 cm/s to 19 cm/s in arterial while ranges from 1.5 cm/s to 7.1 cm/s in the vein.

The model we establish is a multiphysics model, which contains temperature coupling and flow coupling. About the boundary, the Marangoni effect works here [49]. The boundary condition can be described as the following equation:

In the simulation, the fine degree of the calculation is reasonably considered, which is the model meshing. In the area of the blood vessels, we mesh it with the finest gridding. The other parts are meshed more roughly compared to the blood vessels.

With a series of parameters we set in simulation, we obtain the empirical data (the temperature corresponding to the blood vessel velocity). And with the empirical data, the relationship between the temperature and the blood flow velocity can be characterized as a polynomial. And once we get the polynomial and the temperature observed in reality, the blood flow velocity can be described.

3. Simulations and Results

The blood vessels in human fingers are diverse, and they hold comparatively great differences on the diameter (800 μm∼1.8 mm). Hence, we choose a suitable range (0.3 mm∼0.45 mm) to describe the blood vessel which is built in the model. In simulation, the physical dimension of model is shown in Figure 6. On the basis of actual size, the radiuses of the three cylinders (the blood vessel, the finger bone, and the geometric shape of finger) are set as 0.4 mm, 0.65 mm, and 0.85 mm, respectively. In the meantime, the lengths of these three cylinders are all set as 80 mm.

Based on the operation above, we calculate the model by using gridding. From a cross-sectional cutaway view of one certain simulation, we can visually observe the whole energy flowing process, as shown in Figure 7. Figure 7(a) shows the initial state, heat energy still gathered at the source. From Figures 7(b) to 7(e), we can visually observe the heat diffusion in the finger. Due to the blood flow, the heat distribution is not symmetrical. Comparing to the Figure 7(e), the warmer scope of the Figure 7(f) is slightly smaller.

In this simulation, a curve is fitted by using multiple simulation results. Within the reasonable range (3 cm/s∼12.6 cm/s) of the blood flow velocity, we conduct 51 times of simulation to get 51 pairs of results (“velocity-temperature”), where the relevant factors (the blood vessel number, the blood flow direction, and the blood vessel radius) are temporarily fixed. Nevertheless, we aim not only at these 51 pairs of results, but the influence of these 3 factors. Therefore, we change the value of these factors for another several simulations again. Eventually, we obtain the integrated empirical curves which are shown below. Figures 8 and 9 show the experience curves of blood flow velocities and temperatures, and these curves signally reflect the difference between each kind of simulation data, where the data come from several different conditions.

The differences between these empirical curves are discussed, which come from the 3 factors (the blood vessel number, the blood vessel radius, and the blood flow direction).

According to Figure 8, when only one blood vessel is considered, the larger the blood vessel is, the more heat is taken away, and the empirical curve shifts down more. In the meantime, the slopes of these curves are approximative.

According to Figure 9, when the radius of blood vessels is 0.3 mm and the blood flow directions are the same, more blood vessels are considered, more heat is taken away, and the empirical curve shifts down more. In the meantime, the slopes of these curves are approximative.

The other two curves in Figure 9 describe the situations that the blood flow directions are reverse. The curve marked as “2r” is upon the curve marked as “4r,” and both the two curves decline when blood flow velocity increases, meanwhile, the degrees of inclination of these two curves are similar. Obviously, these two curves indicate that more blood vessels take more heat away. And a higher blood flow velocity brings more heat away too. When the number of blood vessel is the same, the curve (2r, 4r) under the same blood flow direction is sharper than the curve (2s, 4s) under the reverse blood flow direction.

In a certain condition, a polynomial can be provided via fitting the corresponding empirical curve, where the relationship between the temperature and the blood velocity in such a condition is described.

Then, the blood flow velocity can be obtained once the temperature and the corresponding polynomial are established. For instance, when the blood vessel’s radius is set to 0.4 mm and only one blood vessel is built in the model, the several pairs of data that are obtained from the simulation are marked as blue points, which are shown in Figure 10. Besides, relying on these data, the fitting curve is marked with red color and shown below. For better fitting performance, the error precision is controlled within 0.00001. Afterwards, the corresponding polynomial is provided and shown in the following equation:

4. Conclusion

In this paper, a noninvasive method has been presented for the blood flow velocity measurement. Compared with the Doppler based method, the structure of the method in this paper is simpler. Besides, the ring that came up in the method is smaller than that used in the Doppler anemometry in the market, so the device of this method is easier to carry, and the work sites of this method are more flexible.

By using this method in modeling, we have obtained a schematic diagram of heat transfer process and 10 empirical curves. Relying on these empirical curves, one of the relationships between the blood velocity and the observed temperature has been described with a polynomial. Moreover, the influences of the 3 factors (the blood vessel number, the blood vessel radius, and the blood flow direction) on the empirical curves have been discussed.

Even though the parameters of the model are set according to actual data, the polynomial in this paper is not precise enough due to individual differences. The purpose of this paper is to analyze the relationship according to a relatively primitive model. Refining and optimizing are needed for a more feasible method in later study. For instance, the temperature of the heat source is stable in the primitive model, and different states of the temperature may lead to a better result. Changing the outline of the finger, ring material, and arrangement of blood vessels can make the model more realistic. The diverse group of people is also an important aspect for improving the method. The primitive model will be improved with more reasonable conditions.

Conflicts of Interest

The authors declare that there are no conflicts of interest regarding the publication of this paper.

Acknowledgments

This research was supported in part by Shenzhen Peacock Plan (KQTD201507161644225).

References

  1. M. A. Gimbrone and G. García-Cardeña, “Vascular endothelium, hemodynamics, and the pathobiology of atherosclerosis,” Cardiovascular Pathology, vol. 22, no. 1, pp. 9–15, 2013. View at: Publisher Site | Google Scholar
  2. M. J. Simmonds, H. J. Meiselman, and O. K. Baskurt, “Blood rheology and aging,” Journal of Geriatric Cardiology, vol. 10, no. 3, pp. 291–301, 2013. View at: Google Scholar
  3. I. A. Tikhomirova, O. O. Anna, and G. M. Svetlana, “Microcirculation and blood rheology in patients with cerebrovascular disorders,” Clinical Hemorheology and Microcirculation, vol. 49, no. 1–4, pp. 295–305, 2011. View at: Google Scholar
  4. J. Li, F. Tang, X. Xu, and D. Wang, “Noninvasive method of measuring blood flow rate by heat diffusion,” Chinese Journal of Scientific Instrument, vol. 27, no. 6, pp. 1178–1180, 2006. View at: Google Scholar
  5. F. Tang, X. Wang, and D. Wang, “Blood flow rate measurements using thermal diffusion,” Chinese Journal of Scientific Instrument, vol. 29, no. 5, pp. 978–981, 2008. View at: Google Scholar
  6. B. Shen, “Measurement of blood microcirculation velocity by image processing,” Chinese Journal of Biomedical Engineering, vol. 24, no. 6, p. 717, 2005. View at: Google Scholar
  7. N. A. Lassen, O. Henriksen, and P. Sejrsen, “Indicator methods for measurement of organ and tissue blood flow,” in Comprehensive Physiology, pp. 21–63, American Physiological Society, Rockville, MD, USA, 2011. View at: Google Scholar
  8. S. Puri, J. K. Park, F. Modersitzki, and D. S. Goldfarb, “Radioisotope blood volume measurement in hemodialysis patients,” Hemodialysis International, vol. 18, no. 2, pp. 406–414, 2014. View at: Publisher Site | Google Scholar
  9. W. Li and A. Ahn, “Effect of acupuncture manipulations at LI4 or LI11 on blood flow and skin temperature,” Journal of Acupuncture and Meridian Studies, vol. 9, no. 3, pp. 128–133, 2016. View at: Publisher Site | Google Scholar
  10. L. Grote, D. Zou, H. Kraiczi, and J. Hedner, “Finger plethysmography—a method for monitoring finger blood flow during sleep disordered breathing,” Respiratory Physiology & Neurobiology, vol. 136, no. 2-3, pp. 141–152, 2003. View at: Publisher Site | Google Scholar
  11. L. M. Blendis, V. C. Roberts, M. Spiro, and R. Williams, “The comparative measurement of splenic blood flow using 133xenon and an electromagnetic flowmeter,” Cardiovascular Research, vol. 4, no. 1, pp. 44–49, 1970. View at: Publisher Site | Google Scholar
  12. J. P. Woodcock, Theory and Practice of Blood Flow Measurement, Butterworth-Heinemann, Oxford, UK, 2013.
  13. R. Salo, “Method and apparatus for monitoring left ventricular work or power,” US Patent 6,892,095, 2005. View at: Google Scholar
  14. D. Zikich and D. Zikic, “Arterial blood flow sensor,” in Proceedings of 4th European Conference of the International Federation for Medical and Biological Engineering, vol. 22, pp. 1158–1162, Springer, Berlin, Germany, 2009. View at: Google Scholar
  15. V. Poirier, “Blood flow meter,” US Patent 8715151, 2014. View at: Google Scholar
  16. W. Nichols, M. O’Rourke, and C. Vlachopoulos, McDonald’s Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles, CRC press, Boca Raton, FL, USA, 2011.
  17. X. Huang, J. Guo, X. Wang, M. Yan, Y. Kang, and H. Yu, “A contact-imaging based microfluidic cytometer with machine-learning for single-frame super-resolution processing,” PLoS One, vol. 9, no. 8, Article ID e104539, 2014. View at: Publisher Site | Google Scholar
  18. P. Xue, K. Ye, J. Gao et al., “Isolation and elution of Hep3B circulating tumor cells using a dual-functional herringbone chip,” Microfluidics and Nanofluidics, vol. 16, no. 3, pp. 605–612, 2014. View at: Publisher Site | Google Scholar
  19. J. Guo, H. Li, Y. Chen, and Y. Kang, “A microfluidic impedance cytometer on printed circuit board for low cost diagnosis,” IEEE Sensors Journal, vol. 14, no. 7, pp. 2112–2117, 2014. View at: Publisher Site | Google Scholar
  20. J. Guo, C. M. Li, and Y. Kang, “PDMS-film coated on PCB for AC impedance sensing of biological cells,” Biomedical Microdevices, vol. 16, no. 5, pp. 681–686, 2014. View at: Publisher Site | Google Scholar
  21. J. Guo, W. Lei, X. Ma, P. Xue, Y. Chen, and Y. Kang, “Design of a fluidic circuit-based microcytometer for circulating tumor cell detection and enumeration,” IEEE Transactions on Biomedical Circuits and Systems, vol. 8, no. 1, pp. 35–41, 2014. View at: Publisher Site | Google Scholar
  22. J. Guo, T. S. Pui, Y. L. Ban, A. R. Abdur Rahman, and Y. Kang, “Electrokinetic analysis of cell translocation in low-cost microfluidic cytometry for tumor cell detection and enumeration,” IEEE Transactions on Biomedical Engineering, vol. 60, no. 12, pp. 3269–3275, 2013. View at: Publisher Site | Google Scholar
  23. J. Guo, T. S. Pui, A. R. A. Rahman, and Y. Kang, “3D numerical simulation of a coulter counter array with analysis of electrokinetic forces,” Electrophoresis, vol. 34, no. 3, pp. 417–424, 2013. View at: Publisher Site | Google Scholar
  24. J. Guo, L. Chen, Y. Ai et al., “Numerical and experimental characterization of solid-state micropore-based cytometer for detection and enumeration of biological cells,” Electrophoresis, vol. 36, no. 5, pp. 737–743, 2015. View at: Publisher Site | Google Scholar
  25. J. Guo, X. Huang, D. Shi et al., “Portable resistive pulse-activated lens-free cell imaging system,” RSC Advances, vol. 4, no. 99, pp. 56342–56345, 2014. View at: Publisher Site | Google Scholar
  26. J. Guo, J. L. W. Li, Y. Chen, L. Y. Yeo, J. R. Friend, and Y. Kang, “RF-activated standing surface acoustic wave for on-chip particle manipulation,” IEEE Transactions on Microwave Theory and Techniques, vol. 62, no. 9, pp. 1898–1904, 2014. View at: Publisher Site | Google Scholar
  27. J. Guo, X. Liu, K. Kang, Y. Ai, Z. Wang, and Y. Kang, “A compact optofluidic cytometer for detection and enumeration of tumor cells,” Journal of Lightwave Technology, vol. 33, no. 16, pp. 3433–3438, 2015. View at: Publisher Site | Google Scholar
  28. J. Guo, X. Ma, N. V. Menon et al., “Dual fluorescence-activated study of tumor cell apoptosis by an optofluidic system,” IEEE Journal of Selected Topics in Quantum Electronics, vol. 21, no. 4, pp. 392–398, 2015. View at: Google Scholar
  29. J. Guo, “Uric acid monitoring with a smartphone as the electrochemical analyzer,” Analytical Chemistry, vol. 88, no. 24, pp. 11986–11989, 2016. View at: Publisher Site | Google Scholar
  30. J. Guo, X. Huang, and Y. Ai, “On-demand lensless single cell imaging activated by differential resistive pulse sensing,” Analytical Chemistry, vol. 87, no. 13, pp. 6516–6519, 2015. View at: Publisher Site | Google Scholar
  31. J. Guo and X. Ma, “Simultaneous monitoring of glucose and uric acid on a single test strip with dual channels,” Biosensors and Bioelectronics, vol. 94, pp. 415–419, 2017. View at: Publisher Site | Google Scholar
  32. J. Guo, “Smartphone-powered electrochemical dongle for point-of-care monitoring of blood beta-ketone,” Analytical Chemistry, vol. 89, no. 17, pp. 8609–8613, 2017. View at: Publisher Site | Google Scholar
  33. K. K. Shung, Diagnostic Ultrasound: Imaging and Blood Flow Measurements, CRC press, Boca Raton, FL, USA, 2015.
  34. L. Qi, J. Zhu, A. M. Hancock et al., “Fully distributed absolute blood flow velocity measurement for middle cerebral arteries using doppler optical coherence tomography,” Biomedical Optics Express, vol. 7, no. 2, pp. 601–615, 2016. View at: Publisher Site | Google Scholar
  35. H. C. Choo, K. Nosaka, J. J. Peiffer, M. Ihsan, C. C. Yeo, and C. R. Abbiss, “Reliability of laser doppler, near-infrared spectroscopy and doppler ultrasound for peripheral blood flow measurements during and after exercise in the heat,” Journal of Sports Sciences, vol. 35, no. 17, pp. 1715–1723, 2017. View at: Publisher Site | Google Scholar
  36. H. Münstedt, “Investigations of slip in capillary flow by laser-doppler velocimetry and their relations to melt fracture,” AIP Conference Proceedings, vol. 1662, no. 1, Article ID 030001, 2015. View at: Publisher Site | Google Scholar
  37. A. P. Shepherd and P. Å. Öberg, Laser-Doppler Blood Flowmetry, vol. 107, Springer Science & Business Media, Berlin, Germany, 2013.
  38. L. Fu and S. Cai, Blood Density Measurement and Its Application in Basic and Clinical Medicine, vol. 24, China Academic Journal Electronic Publishing House, Minneapolis, MN, USA, 2001.
  39. G. Zhang, “Influence and correlation of exercise on bone mineral density and body composition of college students,” China Sport Science and Technology, vol. 44, no. 5, pp. 56–62, 2008. View at: Google Scholar
  40. K. Yaffe, J. Nettiksimmons, R. Murphy et al., “Adipose tissue density and cognitive decline and dementia,” Alzheimer’s & Dementia, vol. 11, no. 7, pp. P827–P828, 2015. View at: Publisher Site | Google Scholar
  41. J. W. Otte, M. A. Merrick, C. D. Ingersoll et al., “Subcutaneous adipose tissue thickness alters cooling time during cryotherapy,” Archives of Physical Medicine and Rehabilitation, vol. 83, no. 11, pp. 1501–1505, 2002. View at: Publisher Site | Google Scholar
  42. D. B. Rodrigues, P. F. Maccarini, S. Salahi et al., “Numerical 3D modeling of heat transfer in human tissues for microwave radiometry monitoring of brown fat metabolism,” in Proceedings of Energy-based Treatment of Tissue and Assessment VII (SPIE BiOS), vol. 8584, San Francisco, CA, USA, February 2013. View at: Publisher Site | Google Scholar
  43. M. Klarhöfer, B. Csapo, C. Balassy, J. C. Szeles, and E. Moser, “High‐resolution blood flow velocity measurements in the human finger,” Magnetic Resonance in Medicine, vol. 45, no. 4, pp. 716–719, 2001. View at: Publisher Site | Google Scholar
  44. K. C. Toh, X. Y. Chen, and J. C. Chai, “Numerical computation of fluid flow and heat transfer in microchannels,” International Journal of Heat and Mass Transfer, vol. 45, no. 26, pp. 5133–5141, 2002. View at: Publisher Site | Google Scholar
  45. W. Zhang, J. R. Xu, Q. Lu, S. Ye, and X. S. Liu, “High-resolution magnetic resonance angiography of digital arteries in SSc patients on 3 tesla: preliminary study,” Rheumatology, vol. 50, no. 9, pp. 1712–1719, 2011. View at: Publisher Site | Google Scholar
  46. H. M. Schmidt and U. Lanz, Surgical Anatomy of the Hand, Thieme, Stuttgart, Germany, 2003.
  47. B. K. Berkovitz, G. R. Holland, and B. J. Moxham, Oral Anatomy, Histology and Embryology, Elsevier, New York City, NY, USA, 2016.
  48. M. Gradisar and L. Lack, “Relationships between the circadian rhythms of finger temperature, core temperature, sleep latency, and subjective sleepiness,” Journal of Biological Rhythms, vol. 19, no. 2, pp. 157–163, 2004. View at: Publisher Site | Google Scholar
  49. A. B. Comsol, COMSOL Multiphysics User’s Guide, vol. 333, COMSOL, Burlington, MA, USA, 2005.
  50. J. R. Welty, C. E. Wicks, G. Rorrer et al., Fundamentals of Momentum, Heat, and Mass Transfer, John Wiley & Sons, Hoboken, NJ, USA, 2009.
  51. Y. He, H. Liu, R. Himeno, J. Sunaga, N. Kakusho, and H. Yokota, “Finite element analysis of blood flow and heat transfer in an image-based human finger,” Computers in Biology and Medicine, vol. 38, no. 5, pp. 555–562, 2008. View at: Publisher Site | Google Scholar
  52. T. L. Bergman and F. P. Incropera, Fundamentals of Heat and Mass Transfer, John Wiley & Sons, Hoboken, NJ, USA, 2011.
  53. F. Liu, K. C. Pawan, G. Zhang, and J. Zhe, “Microfluidic magnetic bead assay for cell detection,” Analytical Chemistry, vol. 88, no. 1, pp. 711–717, 2015. View at: Publisher Site | Google Scholar
  54. G. K. Batchelor, An Introduction to Fluid Dynamics, Cambridge University Press, Cambridge, UK, 2000.
  55. R. L. Panton, Incompressible Flow, John Wiley & Sons, Hoboken, NJ, USA, 2006.
  56. N. Riley, “Steady streaming,” Annual Review of Fluid Mechanics, vol. 33, no. 1, pp. 43–65, 2001. View at: Publisher Site | Google Scholar
  57. I. Marusic, B. J. McKeon, P. A. Monkewitz, H. M. Nagib, A. J. Smits, and K. R. Sreenivasan, “Wall-bounded turbulent flows at high reynolds numbers: recent advances and key issues,” Physics of Fluids, vol. 22, no. 6, Article ID 65103, 2010. View at: Publisher Site | Google Scholar

Copyright © 2018 Fu Yusheng et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


More related articles

1102 Views | 329 Downloads | 0 Citations
 PDF  Download Citation  Citation
 Download other formatsMore
 Order printed copiesOrder

Related articles

We are committed to sharing findings related to COVID-19 as quickly and safely as possible. Any author submitting a COVID-19 paper should notify us at help@hindawi.com to ensure their research is fast-tracked and made available on a preprint server as soon as possible. We will be providing unlimited waivers of publication charges for accepted articles related to COVID-19. Sign up here as a reviewer to help fast-track new submissions.