Analysis of a Pulse Rate Variability Measurement Using a Smartphone Camera

Table 3

Comparison values for the correspondence values among PRV and HRV indices, when comparing PPG to ECG, and the channels of ECG. Here, PC stands for the Pearson correlation coefficient (with values smaller than 10−10), m and b represent the coefficients for the linear regression, and bias, SD, and BAR values are for the Bland-Altman analysis. The worst values are in bold.

ECG-PPGECG(1)-ECG(2)ECG(1)-ECG(3)ECG(2)-ECG(3)

Mean RR (ms)PC1111
m1111
b−0.019−0.0250.0280.032
Bias−0.0210.00150.007−0.003
SD0.00930.0210.0610.026
BAR<0.001<0.001<0.001<0.001

RMSSD (ms)PC0.99610.9980.999
m0.9981.0040.990.99
b2.53−0.250.670.87
Bias2.64−0.130.430.46
SD1.790.441.211.01
BAR0.1040.0270.0760.07

SDNN (ms)PC0.998111
m1.0051.0021.0010.998
b1.055−0.1130.0750.162
Bias1.27−0.0200.1360.092
SD0.7760.1480.4640.260
BAR0.0350.00680.0210.012

pNN50 (%)PC0.9930.9990.9950.998
m1.070.9880.9930.997
b1.670.0760.3050.122
Bias2.67−0.0830.2210.094
SD2.430.8691.6310.918
BAR0.3060.1250.2500.168

TP (total power, ms2)PC0.998111
m1.041.0031.0091.005
b16.523.01−2.49−2.032
Bias50.15−0.4845.2611.788
SD46.158.76821.8413.83
BAR0.100.0190.0480.0335

LP (low-frequency power, ms2)PC1111
m1.0511.0041
b−1.140.260.1670.307
Bias15.960.161.500.243
SD16.032.265.742.43
BAR0.0890.0120.0330.015

HP (high-frequency power, ms2)PC0.99510.9990.999
m1.061.0021.0171.013
b18.514−0.94−0.5861.18
Bias33.87−0.553.681.49
SD33.187.7116.4111.91
BAR0.2470.0640.130.11