Research Article

Modelling and Experiment Based on a Navigation System for a Cranio-Maxillofacial Surgical Robot

Table 1

The results of the registration accuracy test.

The target pointThe registration error (mm)
Group oneGroup twoGroup threeGroup fourGroup fiveGroup sixMean ± SD

Right subtemporal0.700.670.630.660.650.450.63 ± 0.08
Right foramen ovale0.660.780.840.730.750.520.71 ± 0.10
External orifice of the right carotid artery0.420.620.500.550.630.720.57 ± 0.10
Right jugular foramen0.540.760.820.680.690.620.69 ± 0.09
Right styloid process0.350.930.410.720.770.550.62 ± 0.20
Left subtemporal0.510.860.730.350.590.770.64 ± 0.17
Left foramen ovale0.680.780.690.710.870.680.74 ± 0.07
External orifice of the left carotid artery0.720.790.710.620.650.380.65 ± 0.13
Left jugular foramen0.540.660.950.450.920.810.72 ± 0.19
Left styloid process0.720.570.630.540.590.920.66 ± 0.13
Mean ± SD0.58 ± 0.130.74 ± 0.110.69 ± 0.150.60 ± 0.120.71 ± 0.110.64 ± 0.16