Research Article
Modelling and Experiment Based on a Navigation System for a Cranio-Maxillofacial Surgical Robot
Table 1
The results of the registration accuracy test.
| The target point | The registration error (mm) | Group one | Group two | Group three | Group four | Group five | Group six | Mean ± SD |
| Right subtemporal | 0.70 | 0.67 | 0.63 | 0.66 | 0.65 | 0.45 | 0.63 ± 0.08 | Right foramen ovale | 0.66 | 0.78 | 0.84 | 0.73 | 0.75 | 0.52 | 0.71 ± 0.10 | External orifice of the right carotid artery | 0.42 | 0.62 | 0.50 | 0.55 | 0.63 | 0.72 | 0.57 ± 0.10 | Right jugular foramen | 0.54 | 0.76 | 0.82 | 0.68 | 0.69 | 0.62 | 0.69 ± 0.09 | Right styloid process | 0.35 | 0.93 | 0.41 | 0.72 | 0.77 | 0.55 | 0.62 ± 0.20 | Left subtemporal | 0.51 | 0.86 | 0.73 | 0.35 | 0.59 | 0.77 | 0.64 ± 0.17 | Left foramen ovale | 0.68 | 0.78 | 0.69 | 0.71 | 0.87 | 0.68 | 0.74 ± 0.07 | External orifice of the left carotid artery | 0.72 | 0.79 | 0.71 | 0.62 | 0.65 | 0.38 | 0.65 ± 0.13 | Left jugular foramen | 0.54 | 0.66 | 0.95 | 0.45 | 0.92 | 0.81 | 0.72 ± 0.19 | Left styloid process | 0.72 | 0.57 | 0.63 | 0.54 | 0.59 | 0.92 | 0.66 ± 0.13 | Mean ± SD | 0.58 ± 0.13 | 0.74 ± 0.11 | 0.69 ± 0.15 | 0.60 ± 0.12 | 0.71 ± 0.11 | 0.64 ± 0.16 | — |
|
|