Research Article

Real-Time Chest Compression Quality Measurements by Smartphone Camera

Table 2

Detection results for all of the 5 tests included in the experiments.

(cpm) (0->) (%) (0–100) (%)

Smartphone position test ()
RateP1261.3 (0.3)99.7 (0.3)
D1R110P1261.8 (1.3)99.5 (1.2)
30:2P14.5 (3.8)95.9 (3.7)
LightP11.1 (0.3)100 (0)
R100P23.0 (3.4)98.1 (3.7)
R150P211.4 (14.9)89.8 (16.4)
30:2P33.3 (1.4)96.0 (2.1)
R100P31.1 (0.2)99.9 (0.4)

Outdoor test ()
OCBR110P11.7 (0.3)100 (0)
OCR110P11.5 (0.3)100 (0)
OSBR110P11.4 (0.4)99.7 (0.5)
OSR110P11.1 (0.4)100 (0)

Disturbance test ()
D2R110P15.896.0

Random movement test ()
Ran.MovP189.6 (2.5)

CPR summary report test ()
CPRsrR110P1TFSCR6.1 (3.3)
TC2.8 (2.6)
TWC10.0 (9.1)
ACR1.8 (1.2)
NC1.6 (1.0)
Total (all tests)2.7 (5.0)98.0 (5.5)4.5 (3.6)

The results are given in mean Average error, , mean Performance, , and mean Relative error parameter, . Standard deviations are shown in parenthesis. R = rate; P = position; D = disturbance; O = outdoor; B = noisy background; C = cloudy; S = sunny; CPRsr = CPR summary report.