An Efficient Teager Energy Operator-Based Automated QRS Complex Detection

Table 5

Results of QRS detection in MIT-BIH Normal Sinus Rhythm Database (MIT-NSD).

CaseTPFNFPDER%Se%+P%Acc%Time (s)

1626511497100.00999.991100.00099.9911.035
162727992270.11399.97599.91299.8881.3
1627310431200.01999.981100.00099.9811.01
16420106872020.20699.81399.98199.7951.3
1648312157500.04199.959100.00099.9591.02
165399130780.16499.92399.91299.8361.04
167739679100.01099.990100.00099.9901.04
167869510200.02199.979100.00099.9791.029
167951038600ā€”100.000100.000100.0001.31
170528851400.04599.955100.00099.9551.049
1745311258010.009100.00099.99199.9911
1817711907500.04299.958100.00099.9581.32
18184108881310.12999.88199.99199.8721.023
1908812360300.02499.976100.00099.9761.04
19090104816510.63099.38499.99099.3741.33
19093911100ā€”100.000100.000100.0001.028
19140113163900.34599.657100.00099.6571.03
19830148116620.45999.55699.98699.5431.047
Total192452235220.13499.87899.98999.8671.108389