Research Article

ECG Multilead QT Interval Estimation Using Support Vector Machines

Table 2

Validation considering 60 patients acquired with DIGICARDIAC. The validation is performed by comparing with respect to the Cardiosoft estimation.

ā€‰QT Cardiosoft (msec)QT est. (msec)Error (%)

Mean389.53391.622.49
Std. dev.30.3730.941.99
Min.320.00318.400.08
Max.450.00453.308.23