The Impact of Pulsatile Flow on Suspension Force for Hydrodynamically Levitated Blood Pump

1School of Mechanical & Energy Engineering, Zhejiang University of Science & Technology, Hangzhou, China
2State Key Laboratory of Fluid Power & Mechatronic Systems, Zhejiang University, Hangzhou, China
3College of Metrology & Measurement Engineering, China Jiliang University, Hangzhou, China
4School of Mechanical Engineering, Hangzhou Dianzi University, Hangzhou, China

Correspondence should be addressed to Feng Huang; nc.ude.uljc@fh

Received 25 August 2018; Revised 28 March 2019; Accepted 21 May 2019; Published 3 June 2019

Academic Editor: Jinshan Tang

Copyright © 2019 Yang Fu et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Hydrodynamically levitated rotary blood pumps (RBPs) with noncontact bearing are effective to enhance the blood compatibility. The spiral groove bearing (SGB) is one of the key components which offer the suspension force to the RBP. Current studies focus on the suspension performance of the SGB under continuous flow condition. However, the RBP shows pulsatile characteristics in the actual clinical application, which may affect the suspension performance of the SGB. In this paper, the impact of pulsatile flow upon the suspension force from the SGB is studied. A model of the SGB with a groove formed of wedge-shaped spirals is built. Then, the CFD calculation of the hydrodynamic force offered by designed SGB under simulated pulsatile flow is introduced to obtain the pulsatile performance of the suspension force. The proposed method was validated by experiments measuring the hydrodynamic force with different bearing gaps. The results show that the suspension force of the SGB under pulsate flow is the same as under steady flow with equivalent effective pressure. This paper provides a method for suspension performance test of the SGB.

1. Introduction

Implanted rotary blood pumps (RBPs) are considered as replacement therapy for heart disease due to the lack of organ donation. The RBP requires a rotor bearing to maintain longtime operation. Compared with contact bearings, noncontact bearings avoid wearing and thrombus formation around the bearing [1]. Magnetic bearings have been applied in RBPs, enhancing the blood compatibility of the RBPs [2]. However, magnetic bearings need sensors and control system to guarantee the suspension, which increases the complexity of the implanted device. Thus, hydrodynamic bearings, which offer suspension force depending on its hydraulic characteristics, were widely studied for RBPs [35].

Spiral groove bearing (SGB), which is small in size and has excellent loading capacity, is considered as a good structure used for the hydrodynamic bearing. The SGB was first introduced by Kink and Reul in the RBP to minimize the size of the implant device, and the mock-loop test result [6] showed that the hydraulic performance of the RBP was not influenced compared to the contact bearing. Yamane et al. [7] made a new groove shape for the SGB to levitate the impellers of RBP, and thrombus formation was reduced in their design. Experimental studies by Zhu et al. [8] pointed out that the pressure generated by SGBs is sufficient to support the rotor. Han et al. [9, 10] developed a novel SGB that the groove width decreases with increasing spiral radius to improve washout of the RBP. The exposure time of blood flow to high shear stresses of the RBP is demanded to be lower in order to improve its hemocompatibility. Amaral et al. [11] introduced an optimized SGB design which can enhance the washout flow between the rotor and pump casing, significantly decreasing the exposure time and improving the overall efficiency of the pump.

The existing studies of the SGBs are mainly focused on the suspension performance under continuous flow condition, where the RBPs are tested with given constant flow rate and the outlet pressure of zero. However, in the actual clinical application, the inlet pressure and outlet pressure of the RBPs vary according to the periodic pressure change of the ventricular, as the RBPs are connected between the left ventricular and aorta [1214]. Therefore, the flow of the RBPs shows pulsatile characteristics even with fixed rotating speed [12, 13]. Studies [1517] show that the pulsatile characteristics of the RBPs affect the performance of the RBPs, such as hydraulic characteristics, blood compatibility, etc. However, no relevant studies were reported about the impact of the RBPs’ pulsatile characteristics on the suspension performance of the SGBs.

For the conventional SGB with logarithmic spiral grooves, mathematical analysis [18] has been deduced with the conclusion that the hydrodynamic force on the bearing was not affected by the pressure variation. As mentioned in [9], the RBP used novel SGBs with the groove formed of wedge-shaped spirals has superior load/flow characteristics, but it is difficult to construct CFD model of the RBP for analysis of its suspension performance. Thus, computational fluid dynamics (CFD) [19] analyses and experiments were conducted in this article to investigate the behavior of SGBs under physiological pulse conditions. The remainder of this article offers the following:(1)CFD model of the SGB with wedge-shaped spiral grooves is built. Then, the CFD calculation of the hydrodynamic force offered by designed SGB under simulated pulsatile flow is introduced to obtain the pulsatile performance of the suspension force.(2)The proposed method is validated by comparing experimentally measured suspension force with CFD calculation under different bearing gaps.

The impact of pulse flow on the suspension performance of the SGB to the RBP has been studied in this paper, which provides a method for suspension performance test of the SGB. The CFD calculation and experiment test have been implemented. Also, the results show that the suspension force of the SGB under pulsate flow is the same as under steady flow with equivalent effective pressure.

2. Materials and Methods

2.1. Structure of SGB

The structure of the RBP with hydrodynamic bearings is shown in Figure 1. The pump consists of the pump housing, impeller, stator, and shaft. The SGB is fixed on the rotor which offers suspension force to levitate the rotor, whose performance will be discussed. There is a big difference in size between the pump and the groove, where the size order of the groove depth and the bearing gap is  m and the pump as  m. As mentioned in the method part, more than 1,800,000 elements are used to fully represent the computational domain of the SGB. Correspondingly, the number of grids will be very huge to cover the whole blood pump, which will exceed the capability of general computers. Thus, only SGB is modeled and studied in this paper.

Figure 1: The schematic of the RGB.

The wedge-shaped SGB is based on the design of Han et al. [9], where the groove width decreases along the flow path. Thus, the relative motion between the SGB and the corresponding surface facilitates the flow through the grooves. The purpose of the novel SGB structure is to maintain high load-carrying capacity while producing a large flow rate to reduce blood damage. The two groove profiles in polar coordinates are expressed aswhere is the radius, is the groove width, is the phase angle, is the spiral angle and remains constant, is the radius at zero polar angle, and is the groove width at . And the groove width is a function of :where , , and . Thus, the groove width decreases along the radius. The geometry of a SGB is shown in Figure 2. The groove depth and bearing gap are and , respectively. The total distance between the groove and the corresponding surface is .

Figure 2: The axial and cross-sectional view of a SGB.
2.2. Numerical Analysis

Numerical simulation was carried out to calculate the suspension force of the SGB under continuous and pulsatile flow condition. According to Han’s study [9], the geometrical parameters of the SGB used in the simulation are listed in Table 1, with groove depth and the spiral angle of 80 μm and 20°. Also, the number of grooves was set to 8.

Table 1: Geometrical parameters of tested SGB.

As shown in Figure 3, three-dimensional grids of the flow field were developed with ANSYS ICEM CFD. Prism grids are generated using the same settings for all cases, and the boundary-layer grids include more than 5 layers of elements. The total number of elements was between 1,800,000 and 3,000,000 according to the actual fluid domain. A mesh refinement study was conducted for 50,000 up to 10,000,000 elements. The estimated hydrodynamic force showed a deviation lower than 5% for meshes over 1,700,000 elements. The computation was conducted through ANSYS CFX and assumed to achieve convergence when the RMS (root mean square) residuals were below 1e−5. In this paper, simulations were carried under continuous flow and pulsatile flow as follows.

Figure 3: CFD mesh result of the SGB.

First, steady-state simulations were performed to acquire the suspension forced by the SGB. In the normal operation of the SGB in the RBP, the bearing gap is commonly between 20 μm to 300 μm [7, 11, 20]. Therefore, the bearing gap in the simulation was chosen from 20 μm to 300 μm, and the data were calculated every 20 μm. The boundary condition was set with inlet and outlet pressure of zero. The estimated hydrodynamic force was further compared with the experiments.

Second, numerical simulation was operated under pulsatile flow condition to figure out whether the suspension force of the SGB was affected. Shi and Korakianitis [21] built a numerical model of the cardiovascular system, where the left ventricular pressure (Plv) and the aortic pressure (Pao) in healthy condition were given as shown in Figure 4. The left ventricular pressure varies between 10 and 120 mmHg, and the aortic pressure changes from 80 to 120 mmHg. In the actual operation, the inlet and outlet of the RBP were connected to the ventricular and aorta, respectively. Therefore, the pressure at r1 and r2 was set equal to Plv and Pao to simulate the pulsate flow condition, which was close to the actual operation condition of the SGB. For comparison, a series of steady-state simulations were carried out with time invariant boundary conditions based on Pao-Plv at a certain time, as shown in Table 2.

Figure 4: Left ventricular pressure (Plv) and aorta pressure (Pao) with the cardiovascular model in healthy conditions [21].
Table 2: Boundary conditions for different pressure heads.
2.3. Experimental Validation

Experiments were carried out to validate the suspension performance test method and the effectiveness of the simulation presented above. The schematic of the experimental setup is shown in Figure 5, which is modified from previous devices on suspension force test for hydrodynamically levitated blood pump reported in colleagues’ research [11, 22]. A SGB manufactured with ABS was fixed on a motor shaft (EC32 Flat, maxon motor ag, Sachseln, Switzerland) which was controlled by a monitor (maxon motor ag, Sachseln, Switzerland). The motor was supported by a 5 mm lifting platform (Winner Optics, WN01VM5) with 5 μm resolution through a cantilever made of aluminum alloy. The 5 mm lifting platform was mounted on a 60 mm Z stage (Winner Optics, WN08VM60) that regulates height roughly. The SGB to be tested was immerged in a cylindrical reservoir, and the fluid was a mixture of water and glycerol with 38% glycerol in weight. A 6-DOF force and moment sensor (ATI Nano 43) was fixed between the reservoir and a X-Y stage (Winner Optics, WN202WM25M) to adjust the SGB to the center of the reservoir. All signals were acquired by a DAQ card (National Instruments, Austin, Texas, USA), and a control program was written with LabVIEW (National Instruments, Austin, Texas, USA) to monitor the motor speed and the sensor response.

Figure 5: Test system for the measurement of hydrodynamic force.

In this experiment, the hydrodynamic force on the counter plane of the SGB with different bearing gaps changed by the elevator-platform was measured. According to Newton’s third law, the suspension force offered by the SGB was equal to the hydrodynamic force on the counter plane. The geometrical parameters of the SGB were designed the same to the simulation. The rotating speed of the motor was fixed at 3000 rpm, the same as the specified speed of the blood pump. The bearing gap ranged from 80 μm to 300 μm in 20 μm steps.

2.4. Validation of the Model Accuracy

The suspension force of the SGB measured by the experiment was compared with the CFD calculated result to validate the effectiveness of the CFD model, as shown in Figure 6. In the experiment and CFD calculation, the pressure of the inlet and outlet was set to zero.

Figure 6: Comparison between experiment and simulation results under steady state.

The suspension force offered by SGB decreases with the increase of the bearing gap. It was found that the simulation results matched the experiment results well especially when the bearing gap was larger than a critical value, which was about 60 μm in this design. When the bearing gap was small, it was found that the suspension force of the simulation was larger than the experimental value. As shown in Figure 6, the suspension force showed a large discrepancy with the variation of the bearing gap when the bearing gap was below 50 μm. In the experiment, the bearing gap was regulated by the displacement platform, leading to the difficulty in adjusting the bearing gap to a precision level, especially in the tiny distance range. Therefore, the simulation result was different from the experiment when the bearing gap was small. In general, the experiment result fits the simulation. Thus, the conclusion could be made that the numerical simulation model built above was valid to estimate the suspension forced by the SGB.

2.5. Impact of Pulsatile Flow on the Hydrodynamic Force

The hydrodynamic force under pulsatile flow was calculated by CFD, where the pressure at inlet and outlet of the SGB was set equal to Plv and Pao, as shown in Figure 7. First, the transient simulation was carried out with the pressure varied continuously according to the Plv and Pao, and the hydrodynamic force was calculated continuously. Then, CFD calculation was carried out under steady state with several given pressures the same as the transient condition as comparisons. It can be found that the hydrodynamic force under pulsatile flow was the same as steady state value. In this calculation, the hydrodynamic force was calculated including the pressure on the surface of the SGB. Thus, the hydrodynamic force was relative to the set pressure on inlet and outlet. Through comparison between the transient results and steady state results, it can be concluded that the suspension force of the SGB under pulsating flow is the same as under steady flow with equivalent effective pressure.

Figure 7: Comparison between transient and steady-state results by CFD calculation.

3. Conclusion

The impact of the pulsatile flow on the suspension force of the SGB was studied in this paper. A model of the SGB with a groove formed of wedge-shaped spirals was built. CFD calculation of the designed SGB was carried out to obtain the effects of the suspension force according to the changing flow. The hydrodynamic force with different bearing gaps was measured by experiment. The measured suspension forces were in accord with the CFD calculation results under steady state, which validated the effectiveness of CFD calculation for the SGB. Also, the CFD calculation results showed that the suspension force of the SGB under pulsating flow is the same as under steady flow with equivalent effective pressure. In this study, the simulation and experiments were implemented for the SGB, not including the whole region of the RBP. The motion of the pump rotor is affected by the coupling of the suspension force of the SGB, the normal force of the impeller, and the magnetic force of the motor. In future works, the influence of the latter two forces will be considered and studied.

Data Availability

The data used to support the findings of this study are available from the corresponding author upon request.

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Acknowledgments

This work was supported by the National Natural Science Foundation of China (grant no. 51505455), funded by the Open Foundation of the State Key Laboratory of Fluid Power and Mechatronic Systems (nos. GZKF-2018017 and GZKF-201713), and supported by the Zhejiang Provincial Natural Science Foundation (nos. LY18E050007 and LGG18E080005).

References

  1. D. Timms, “A review of clinical ventricular assist devices,” Medical Engineering & Physics, vol. 33, no. 9, pp. 1041–1047, 2011. View at Publisher · View at Google Scholar · View at Scopus
  2. T. Masuzawa, “Magnetically suspended motor system applied to artificial hearts and blood pumps,” Proceedings of the Institution of Mechanical Engineers, Part I: Journal of Systems and Control Engineering, vol. 231, no. 5, pp. 330–338, 2017. View at Publisher · View at Google Scholar · View at Scopus
  3. S. Kishimoto, Y. Takewa, T. Tsukiya et al., “Novel temporary left ventricular assist system with hydrodynamically levitated bearing pump for bridge to decision: initial preclinical assessment in a goat model,” Journal of Artificial Organs, vol. 21, no. 1, pp. 23–30, 2018. View at Publisher · View at Google Scholar · View at Scopus
  4. G. Foster, “Third-generation ventricular assist devices,” in Mechanical Circulatory and Respiratory Support, pp. 151–186, Elsevier, Amsterdam, Netherlands, 2018. View at Google Scholar
  5. M. Nishida, “Artificial hearts-recent progress: republication of the article published in the Japanese Journal of Artificial Organs,” Journal of Artificial Organs, vol. 20, no. 3, pp. 187–193, 2017. View at Publisher · View at Google Scholar · View at Scopus
  6. T. Kink and H. Reul, “Concept for a new hydrodynamic blood bearing for miniature blood pumps,” Artificial Organs, vol. 28, no. 10, pp. 916–920, 2004. View at Publisher · View at Google Scholar · View at Scopus
  7. T. Yamane, O. Maruyama, M. Nishida et al., “Hemocompatibility of a hydrodynamic levitation centrifugal blood pump,” Journal of Artificial Organs, vol. 10, no. 2, pp. 71–76, 2007. View at Publisher · View at Google Scholar · View at Scopus
  8. L. Zhu, Y. Wu, and Y. Luo, “Experimental evaluation of a novel injection suspended impeller for implantable centrifugal blood pump,” International Journal of Applied Electromagnetics and Mechanics, vol. 52, no. 1-2, pp. 525–530, 2016. View at Publisher · View at Google Scholar · View at Scopus
  9. Q. Han, J. Zou, X. Ruan, X. Fu, and H. Yang, “A novel design of spiral groove bearing in a hydrodynamically levitated centrifugal rotary blood pump,” Artificial Organs, vol. 36, no. 8, pp. 739–746, 2012. View at Publisher · View at Google Scholar · View at Scopus
  10. Q. Han, X. Ruan, W. Chen, and X. Fu, “Numerical simulation and experimental research on passive hydrodynamic bearing in a blood pump,” Chinese Journal of Mechanical Engineering, vol. 26, no. 5, pp. 967–973, 2013. View at Publisher · View at Google Scholar · View at Scopus
  11. F. Amaral, S. Groß-Hardt, D. Timms, C. Egger, U. Steinseifer, and T. Schmitz-Rode, “The spiral groove bearing as a mechanism for enhancing the secondary flow in a centrifugal rotary blood pump,” Artificial Organs, vol. 37, no. 10, pp. 866–874, 2013. View at Publisher · View at Google Scholar · View at Scopus
  12. O. Frazier, T. Myers, S. Westaby, and I. Gregoric, “Use of the jarvik 2000 left ventricular assist system as a bridge to heart transplantation or as destination therapy for patients with chronic heart failure,” Annals of Surgery, vol. 237, no. 5, pp. 631–637, 2003. View at Publisher · View at Google Scholar · View at Scopus
  13. K. Yamazaki, S. Saito, S. Kihara, O. Tagusari, and H. Kurosawa, “Completely pulsatile high flow circulatory support with a constant-speed centrifugal blood pump: mechanisms and early clinical observations,” General Thoracic and Cardiovascular Surgery, vol. 55, no. 4, pp. 158–162, 2007. View at Publisher · View at Google Scholar · View at Scopus
  14. E. V. Potapov, M. Loebe, B. A. Nasseri et al., “Pulsatile flow in patients with a novel nonpulsatile implantable ventricular assist device,” Circulation, vol. 102, no. 3, pp. III183–III187, 2000. View at Publisher · View at Google Scholar
  15. L. V. Belyaev, A. B. Ivanchenko, A. V. Zhdanov, and V. V. Morozov, “Mathematical modeling of hemodynamic characteristics of pumps for pulsatile circulatory support systems,” Biomedical Engineering, vol. 49, no. 1, pp. 24–28, 2015. View at Publisher · View at Google Scholar · View at Scopus
  16. J. Di Paolo, J. F. Insfran, E. R. Fries, D. M. Campana, M. E. Berli, and S. Ubal, “A preliminary simulation for the development of an implantable pulsatile blood pump,” Advances in Biomechanics and Applications, vol. 1, no. 2, pp. 127–141, 2014. View at Publisher · View at Google Scholar
  17. A. L. Edwards, P. Fitzmorris, S. V. Pamboukian, J. F. George, C. M. Wilcox, and S. Peter, “Association of pulsatility with gastrointestinal bleeding in a cohort of heartmate II recipients,” ASAIO Journal, vol. 64, no. 4, pp. 472–479, 2018. View at Publisher · View at Google Scholar · View at Scopus
  18. E. A. Muijderman, “Spiral groove bearings,” Industrial Lubrication and Tribology, vol. 17, no. 1, pp. 12–17, 1965. View at Publisher · View at Google Scholar · View at Scopus
  19. Q. Liu, J. Ye, G. Zhang et al., “Study on the metrological performance of a swirlmeter affected by flow regulation with a sleeve valve,” Flow Measurement and Instrumentation, vol. 67, pp. 83–94, 2019. View at Publisher · View at Google Scholar
  20. W.-K. Chan, K.-T. Ooi, and Y.-C. Loh, “Numerical and in vitro investigations of pressure rise in a new hydrodynamic blood bearing,” Artificial Organs, vol. 31, no. 6, pp. 434–440, 2007. View at Publisher · View at Google Scholar · View at Scopus
  21. Y. Shi and T. Korakianitis, “Numerical simulation of cardiovascular dynamics with left heart failure and in-series pulsatile ventricular assist device,” Artificial Organs, vol. 30, no. 12, pp. 929–948, 2006. View at Publisher · View at Google Scholar · View at Scopus
  22. L. J. Leslie, L. J. Marshall, A. Devitt, A. Hilton, and G. D. Tansley, “Cell exclusion in couette flow: evaluation through flow visualization and mechanical forces,” Artificial Organs, vol. 37, no. 3, pp. 267–275, 2013. View at Publisher · View at Google Scholar · View at Scopus