Research Article
An Optimization Study of Estimating Blood Pressure Models Based on Pulse Arrival Time for Continuous Monitoring
Table 5
The estimated BP errors in different BP models.
| Models | Systolic blood pressure (SBP) | Diastolic blood pressure (DBP) | Mean ± SD (mm·Hg) | MAD (mm·Hg) | Mean ± SD (mm·Hg) | MAD (mm·Hg) |
| MK-EE | 0.07 ± 5.87 | 4.46 | −0.13 ± 4.54 | 3.58 | L-MK | −5.89 ± 12.74 | 9.34 | −3.72 ± 6.79 | 5.91 | MK-BH | 0.11 ± 7.53 | 5.48 | −2.10 ± 5.69 | 4.69 | dMK-BH | −0.01 ± 5.90 | 4.55 | 0.04 ± 4.40 | 3.38 | M-M | 1.11 ± 7.51 | 5.57 | −0.23 ± 6.47 | 5.13 | ANSI/AAMI | ≤ 5 mm·Hg | ≤7 mm·Hg | ≤ 5 mm·Hg | ≤7 mm·Hg | SP10 standard | ≤ 8 mm·Hg | ≤ 8 mm·Hg |
|
|