Research Article

Design of a Machine Learning-Assisted Wearable Accelerometer-Based Automated System for Studying the Effect of Dopaminergic Medicine on Gait Characteristics of Parkinson’s Patients

Table 3

Mean value of gait characteristics and average error rate for the left and right legs.

Sl. no.ParametersMean value (3D motion capture)Mean value (algorithm)Mean error rate (%)

Left leg
1Step time (s)0.570.546.94 ± 2.82
2Stride time (s)1.171.134.76 ± 3.55
3Step length (m)0.370.346.35 ± 2.85
4Stride length (m)0.740.716.51 ± 2.92
5Walking speed (m/s)0.640.617.12 ± 2.74

Right leg
1Step time (s)0.540.567.14 ± 2.52
2Stride time (s)1.181.145.25 ± 3.62
3Step length (m)0.370.346.15 ± 2.81
4Stride length (m)0.740.706.35 ± 2.71
5Walking speed (m/s)0.690.666.72 ± 3.14