Toward Respiratory Support of Critically Ill COVID-19 Patients Using Repurposed Kidney Hollow Fiber Membrane Dialysers to Oxygenate the Blood

Table 3

Blood gas measurements at the inlet and outlet of HFMD at 500 ml/min blood flow rates over three runs. A Leoceed-21N HFMD was used for the first two runs and a Leoceed-18N was used for the third run. Measurements are reported as averages over n samples ±1 standard deviation uncertainty.

pO2 (kPa)pCO2 (kPa)HbSat (%)

Run 1 (21N)
 Hb 10.00 ± 0.05 g/dl
 Inlet (n = 2)4.55 ± 0.0311.26 ± 0.0837.35 ± 0.03
 Outlet (n = 2)15.60 ± 0.17.23 ± 0.0594.70 ± 0.07

Run 2 (21N)
 Hb 10.67 ± 0.04 g/dl
 Inlet (n = 3)4.87 ± 0.038.82 ± 0.0543.73 ± 0.03
 Outlet (n = 4)9.70 ± 0.056.92 ± 0.0485.75 ± 0.04

Run 3 (18N)
 Hb 11.20 ± 0.04 g/dl
 Inlet (n = 4)4.20 ± 0.027.43 ± 0.0442.68 ± 0.02
 Outlet (n = 4)10.23 ± 0.054.65 ± 0.0291.08 ± 0.05