Research Article

Study on Model Iterative Reconstruction Algorithm vs. Filter Back Projection Algorithm for Diagnosis of Acute Cerebral Infarction Using CT Images

Table 1

Comparison results of objective values of image quality among patients from the two groups.

GroupReconstruction methodsCT (Hu)SD (Hu)SNRCNR

Group AFBPGray matter56.23 ± 2.457.82 ± 1.439.41 ± 1.530.40 ± 0.09
White matter36.23 ± 2.125.21 ± 0.9810.06 ± 1.32

Group BFBPGray matter78.69 ± 3.1212.43 ± 1.547.83 ± 0.930.21 ± 0.03
White matter57.12 ± 2.569.26 ± 0.987.82 ± 0.72
iDose4Gray matter65.98 ± 3.1210.79 ± 1.878.21 ± 1.100.24 ± 0.02
White matter48.34 ± 2.178.15 ± 0.857.73 ± 0.91
IMR1Gray matter55.43 ± 3.457.24 ± 1.2710.52 ± 2.130.42 ± 0.06
White matter38.63 ± 1.914.83 ± 0.6210.98 ± 1.17
IMR2Gray matter53.24 ± 3.196.39 ± 1.2811.23 ± 2.150.57 ± 0.11
White matter37.03 ± 4.234.97 ± 0.9211.83 ± 1.86
IMR3Gray matter52.34 ± 3.926.82 ± 1.9711.65 ± 2.310.68 ± 0.21
White matter34.61 ± 1.973.94 ± 0.8212.82 ± 1.91

 ZGray matter234.9670.9836.8765.14
White matter267.25235.8775.61

Gray matter<0.001<0.001<0.001<0.001
White matter<0.001<0.001<0.001