Research Article
Establishment of Normal Range for Thromboelastography in Healthy Middle-Aged and Elderly People of Weihai in China
Table 2
Internal quality control analysis of Level 1 for 30 consecutive days.
| | R (min) | K (min) | Angle (deg) | MA (mm) |
| Channel 1 | 1.53 ± 0.08 (5.2%) | 0.80 ± 0.06 (7.5%) | 83.13 ± 0.75 (0.9%) | 52.83 ± 5.06 (9.5%) | Channel 2 | 1.55 ± 0.14 (9.0%) | 0.83 ± 0.05 (6.0%) | 81.9 ± 0.71 (0.8%) | 54.18 ± 5.32 (9.8%) | Channel 3 | 1.58 ± 0.12 (7.5%) | 0.81 ± 0.05 (6.1%) | 82.24 ± 1.30 (1.5%) | 50.16 ± 5.01 (9.9%) | Channel 4 | 1.60 ± 0.09 (5.6%) | 0.80 ± 0.06 (7.5%) | 82.02 ± 2.65 (3.2%) | 51.38 ± 5.10 (9.9%) | Channel 5 | 1.57 ± 0.05 (3.1%) | 0.80 ± 0.03 (3.7%) | 82.83 ± 2.09 (3.2%) | 52.08 ± 5.04 (9.6%) | Channel 6 | 1.59 ± 0.06 (3.7%) | 0.80 ± 0.04 (5.0%) | 82.18 ± 1.48 (1.8%) | 52.88 ± 5.20 (9.8%) | Reference range | 0–4 | 0–2 | 76–88 | 42–62 |
|
|
The number in brackets is the coefficient of variation.
|