Research Article

VEMERS 2.0: Upgrading of an Emergency Use Ventilator from a Single Mandatory Volume Control Mode of Ventilation (VEMERS 1.0) to 8 Modes of Ventilation

Table 4

Patient General Data in VEMERS UC Clinical tests.

DateTesting hoursVolume (ml)Frequency (bp)IE ratioPEEP (cm H2O)Fio2Patient ageGenderPosition

13/07/202013 : 20 to 21 : 2028032.02.08.045 a 6062maleProno
17/07/202011 : 15 to 19 : 1526033.01.84.045 a7070FemaleProno
21/07/202010 : 00 to 18 : 0036020.02.110.035 a 4072FemaleSupina
22/07/20209 : 00 to 17 : 0034028.01.96.045 a 6059FemaleSupina
29/07/202013 : 20 to 20 : 0027529.71.97.950 a 5572FemaleSupina