Research Article

Clinical Comparison of QUANTA Flash dsDNA Chemiluminescent Immunoassay with Four Current Assays for the Detection of Anti-dsDNA Autoantibodies

Table 3

Clinical performance characteristics for anti-dsDNA antibody assays.

QUANTA Flash dsDNAQUANTA Lite dsDNA SCBioPlex 2200 dsDNAImmuLisa dsDNANOVA Lite dsDNA

Manufacturer’s cut-off used, where equivocal results are considered positive35 IU/mL30 IU/mL5 IU/mL50 IU/mLN/A
Sensitivity in SLE%
(95% CI)
39.3 (27.2–52.7)54.1 (40.8–66.9)44.3 (31.5–57.6)26.2 (15.8–39.1)8.2 (2.7–18.1)
Specificity %  
(95% CI)
96.0 (90.1–98.9)91.0 (83.6–95.8)88.0 (80.0–93.6)96.0 (90.1–98.9)100.0 (96.4–100.0)
LR+9.846.013.696.56
LR−0.630.500.630.770.92
Odds ratio15.612.05.98.5N/A
AUC (95% CI)0.79 (0.72–0.87)0.90 (0.86–0.95)0.68 (0.60–0.77)0.61 (0.52–0.71)0.54 (0.51–0.58)
Cut-off used at 94.0% specificity 27.5 IU/mL49.6 IU/mL12 IU/mL41.9 IU/mLN/A
Sensitivity in SLE% 
at 94.0% specificity (95% CI)
52.5 (39.3–65.4)45.9 (33.1–59.2)21.3 (11.9–33.7)27.9 (17.1–40.8)N/A

AUC: area under the curve; LR: likelihood ratio; CI: confidence interval.