Research Article
Clinical Comparison of QUANTA Flash dsDNA Chemiluminescent Immunoassay with Four Current Assays for the Detection of Anti-dsDNA Autoantibodies
Table 3
Clinical performance characteristics for anti-dsDNA antibody assays.
| |
QUANTA Flash dsDNA | QUANTA Lite dsDNA SC | BioPlex 2200 dsDNA | ImmuLisa dsDNA | NOVA Lite dsDNA |
| Manufacturer’s cut-off used, where equivocal results are considered positive | 35 IU/mL | 30 IU/mL | 5 IU/mL | 50 IU/mL | N/A | Sensitivity in SLE% (95% CI) | 39.3 (27.2–52.7) | 54.1 (40.8–66.9) | 44.3 (31.5–57.6) | 26.2 (15.8–39.1) | 8.2 (2.7–18.1) | Specificity % (95% CI) | 96.0 (90.1–98.9) | 91.0 (83.6–95.8) | 88.0 (80.0–93.6) | 96.0 (90.1–98.9) | 100.0 (96.4–100.0) | LR+ | 9.84 | 6.01 | 3.69 | 6.56 | | LR− | 0.63 | 0.50 | 0.63 | 0.77 | 0.92 | Odds ratio | 15.6 | 12.0 | 5.9 | 8.5 | N/A | AUC (95% CI) | 0.79 (0.72–0.87) | 0.90 (0.86–0.95) | 0.68 (0.60–0.77) | 0.61 (0.52–0.71) | 0.54 (0.51–0.58) | Cut-off used at 94.0% specificity | 27.5 IU/mL | 49.6 IU/mL | 12 IU/mL | 41.9 IU/mL | N/A | Sensitivity in SLE% at 94.0% specificity (95% CI) | 52.5 (39.3–65.4) | 45.9 (33.1–59.2) | 21.3 (11.9–33.7) | 27.9 (17.1–40.8) | N/A |
|
|
AUC: area under the curve; LR: likelihood ratio; CI: confidence interval.
|