Research Article

Development of Monoclonal Antibody-Based EIA for Tetranor-PGDM which Reflects PGD2 Production in the Body

Table 2

Analytical recovery of tetranor-PGDM in artificial urine samples with different tetranor-PGDM concentrations.

Concentration (ng/mL)Mean recoveries (%)
SampleSpikedTheoreticalMeasured

AU-1 ()00.85
0.861.321.49113.5
1.872.102.09100.0
3.913.673.81103.8
9.127.687.0493.0
AU-2 ()05.95
0.865.255.53105.6
1.876.036.53108.4
3.917.606.9991.9
9.1211.629.5382.3

Concentration of tetranor-PGDM was 0.85 (AU-1) and 5.95 (AU-2). The theoretical concentrations were corrected using the recovery value from the SPE procedure (77.1% by LC-MS/MS).