Temporal Trends in X-Ray Exposure during Coronary Angiography and Percutaneous Coronary Intervention

Figure 5

Trends in operator exposure. The ratio between received operator dose and given dose (mSv/DAP) is presented to assess the effects of operator shielding and is not affected by procedure numbers or irradiation time. (a) Pooled trends for mSv/DAP (in red-left, y-axis) and mSv/procedure (blue-right, y-axis) for all operators and fellows working in the cath lab. Between 2013 and 2019, there was a change in mSv/DAP from 4.48 × 10−6 mSv/μGy·m2 to 1.98 × 10−6 mSv/μGy·m2, which corresponds to a 56% reduction (). The reduction in mSv/procedure was larger at 70% () as it is influenced by reduced given dose (DAP) per procedure. (b) Interoperator variability between consultants employed throughout the period.