Research Article

Artificial Intelligence Enabled Fully Automated CMR Function Quantification for Optimized Risk Stratification in Patients Undergoing Transcatheter Aortic Valve Replacement

Table 3

Agreement between manual and automated uncorrected analyses.

Left ventricleBias95% LOAICC (95% CI)COV (%)

LV mass (g)−10.08−84.2 to 64.10.890 (0.846–0.921)21.5
LV EDV (ml)11.13−23.5 to 45.80.978 (0.969–0.984)11.2
LV ESV (ml)4.63−22.6 to 31.90.983 (0.977–0.988)19.5
LV SV (ml)6.69−17.6 to 31.00.935 (0.909–0.954)14.4
LVEF (%)0−12.1 to 12.10.964 (0.950–0.975)10.5

Right ventricleBias95% LOAICC (95% IC)COV (%)
RV EDV (ml)−3.44−40.7 to 33.80.954 (0.936–0.967)13.6
RV ESV (ml)1.37−26.6 to 29.40.955 (0.938–0.968)21.0
RV SV (ml)−4.26−37.1 to 28.60.832 (0.765–0.880)23.3
RVEF (%)−2.44−21.7 to 16.90.804 (0.725–0.860)18.6

EDV: end-diastolic volume; ESV: end-systolic volume; LV: left ventricular; LVEF: left ventricular ejection fraction; RV: right ventricular; RVEF: right ventricular ejection fraction; SV: stroke volume.