Evaluation of Long-Term Inflammatory Responses after Implantation of a Novel Fully Bioabsorbable Scaffold Composed of Poly-L-lactic Acid and Amorphous Calcium Phosphate Nanoparticles
Table 5
Immunohistochemistry results.
1 month
6 months
12 months
24 months
MMP-9 positive index
PLLA
8.65 ± 2.65
13.04 ± 4.77
21.13 ± 4.89
29.32 ± 5.89
PLLA/ACP
7.43 ± 2.71
6.94 ± 3.30
9.58 ± 5.58
8.49 ± 4.23
values
0.18
0.00
0.00
0.00
NF-κB positive index
PLLA
29.38 ± 3.96
34.44 ± 4.07
40.00 ± 3.19
50.62 ± 5.12
PLLA/ACP
26.51 ± 5.15
30.91 ± 3.42
26.24 ± 4.52
20.29 ± 5.39
values
0.07
0.02
0.00
0.00
Notes: Data are presented as mean ± standard deviation. values: the PLLA/ACP group compared with the PLLA group.