| Analyte | Electrode | Limit of detection (ppb) | Reference |
| Carbofuran | AChE/Nafion/BSA/CoPc-SPE | 0.108 | [13] | AChE-TCNQ/SPE | 1.1 | [11] | AuNP/AChE/Au | 7.26 | [14] | PPy-AChE-Geltn-Glut/Pt | 0.12 | [15] | Biocomposite 16% graphite | 0.25 | This paper | Biocomposite 17% graphite | 0.27 | This paper |
| Paraoxon | AChE/Fe3+-BSA-Nafion | 10 | [16] | AChE/PAN-AuNPs/Pt | 0.739 | [17] | AChE/Geltn-Cellulose/SPE | 7.975 | [18] | AChE-Carbon Paste/Cu | 0.86 | [19] | PPy-AChE-Geltn-Glut/Pt | 1.1 | [15] | Biocomposite 16% graphite | 0.25 | This paper | Biocomposite 17% graphite | 0.26 | This paper |
| Dichlorvos | AChE/(MWCNTs/ALB)n/GCE | 61.87 | [5] | Biocomposite 16% graphite | 0.28 | This paper | Biocomposite 17% graphite | 0.26 | This paper |
| Carbaryl | AChE/TMOS/graphite | 2.01 | [3] | Biocomposite 16% graphite | 0.23 | This paper | Biocomposite 17% graphite | 0.24 | This paper |
| Malathion | AChE/TMOS/carbon paste | 58 | [3] | Biocomposite 16% graphite | 0.25 | This paper | Biocomposite 17% graphite | 0.24 | This paper |
|
|