Research Article

Electrochemical Determination of Paracetamol Using Fe3O4/Reduced Graphene-Oxide-Based Electrode

Table 3

Results obtained in the determination of paracetamol in pharmaceutical formulations using the DP-ASV and HPLC.

SampleTablet label
value (mg)
HPLC value  
(mg)
DP-ASV value  
(mg)
Rev.Re

Panadol Extra500502.75496.8695.21.2
Tiffy Dey500494.50505.63103.8–2.3
pms-Mexcold500501.20509.37107.0–1.6

of 3 measurements. Re(%) = 100 × (DP-ASV value − HPLC value)/HPLC value.