Research Article
Simple Label-Free Electrochemical Immunosensor in a Microchamber for Detecting Newcastle Disease Virus
Table 1
The crucial parameters of the two types of electrochemical label-free immunosensors.
| Parameter | PrA/GA immunosensor | SAM/NHS immunosensor |
| Linear range of concentration | 102-106 EID50/mL | 102-106 EID50/mL | Correlation coefficient () | 0.983 | 0.991 | Slope () | 0.0289 | 0.0588 | Standard deviation of the slope (SDS) | 0.00188 | 0.00286 | -intercept () | -0.0099 | 0.2606 | Standard deviation of the -intercept () | 0.0080 | 0.0119 | LOD | 0.95 (9 EID50/mL) | 0.67 (5 EID50/mL) | LOQ | 2.88 (102.88 EID50/mL) | 2.03 (102.03 EID50/mL) |
|
|