Evaluation of Body Composition in Hemodialysis Thai Patients: Comparison between Two Models of Bioelectrical Impedance Analyzer and Dual-Energy X-Ray Absorptiometry
Table 2
Body composition parameters of the hemodialysis patients with BIA S10, BIA 720, and DXA (mean SD).
DXA
BIA S10
BIA 720
Fat mass (kg)
Male
19.28 ± 7.30a
19.16 ± 8.29b
16.68 ± 9.07a,b
Female
20.33 ± 8.94a
19.88 ± 11.32b
16.40 ± 11.28a,b
Total
19.32 ± 7.94a
18.91 ± 9.36b
16.82 ± 9.46a,b
Percentage of fat mass (%)
Male
27.24 ± 6.48a
27.53 ± 10.00b
23.68 ± 10.20a,b
Female
34.11 ± 8.70a
33.68 ± 11.22b
29.30 ± 11.57a,b
Total
29.63 ± 7.98a
29.67 ± 10.77b
25.49 ± 10.90a,b
Fat-free mass (kg)
Male
47.56 ± 5.88
46.46 ± 7.35a
48.88 ± 7.24a
Female
36.66 ± 5.01
35.35 ± 3.96a
37.11 ± 4.35a
Total
43.77 ± 7.63c,d
42.6 ± 8.29a,c
45.08 ± 8.48a,d
Fat mass index (kg/m2)
Male
6.83 ± 2.80a
6.72 ± 3.35b
6.05 ± 3.45a,b
Female
8.37 ± 3.91a
8.27 ± 4.75b
7.30 ± 4.77a,b
Total
7.37 ± 3.29a
7.26 ± 3.93b
6.46 ± 3.93a,b
Fat-free mass index (kg/m2)
Male
17.18 ± 1.88
16.73 ± 1.95a
17.63 ± 1.70a
Female
15.10 ± 2.42
14.50 ± 1.52a
15.19 ± 1.80a
Total
16.46 ± 2.29e
15.95 ± 2.10a,e
16.84 ± 2.07a
BIA = bioelectrical impedance analysis; DXA = dual-energy X-ray absorptiometry; a,bsignificant difference between tools of measurement at in each line; csignificant difference between tools of measurement at in each line; dsignificant difference between tools of measurement at in each line; esignificant difference between tools of measurement at in each line.