Comparison of Optical versus Ultrasonic Biometry in Keratoconic Eyes

Table 2

Pairwise comparison of axial length, anterior chamber depth, lens thickness, and central corneal thickness in keratoconic eyes using the Lenstar and USG biometry.

Pairwise comparison Difference ICC95% limits of agreement value
Mean (SD)Median

CCT Lenstar-US biometer (µm)42−5.4 ± 19.6−4.00.9033.0 to −43.9<0.001*
ACD Lenstar-US biometer (mm)410.18 ± 0.170.170.850.51 to −0.16<0.001*
LT Lenstar-US biometer (mm)41−0.13 ± 0.17−0.140.670.20 to −0.46<0.001*
AL Lenstar-US biometer (mm)42−0.10 ± 0.76−0.160.751.39 to −1.59<0.001*

Paired samples test, ICC: interclass correlation coefficient.
SD: standard deviation, ICC: interclass correlation coefficient, CCT: central corneal thickness, US: ultrasound, ACD: anterior chamber depth, LT: lens thickness, and AL: axial length.