Research Article

Effects of Misalignments in the Retinal Nerve Fiber Layer Thickness Measurements with Spectral Domain Optical Coherence Tomography

Table 2

Comparison of mean ± standard deviation (95% confidence interval) SD-OCT parameters between eyes with and without misalignments in any of the 4 sectors.

SD-OCTAE ()MAE () value

AT (µm)88.0 ± 14.3 (CI 59.9–116.0)87.5 ± 15.3 (CI 57.5–117.5)0.407
Quadrant, (µm)
 Superior 109.1 ± 24.2 (CI 61.6–156.5)106.9 ± 27.8 (CI 52.4–161.4)0.197
 Nasal 70.5 ± 10.3 (CI 50.3–90.6)70.0 ± 12.6 (CI 43.3–94.7)0.607
 Inferior109.1 ± 24.6 (CI 60.8–157.3)109.6 ± 25.1 (CI 60.4–158.8)0.712
 Temporal 62.9 ± 10.9 (CI 41.5–84.2)63.4 ± 12.1 (CI 39.7–87.1)0.339
Clock-hour, (µm)
 1 97.9 ± 26.0 (CI 46.9–148.8)96.4 ± 30.8 (CI 36.0–156.8)0.481
 2 82.0 ± 14.5 (CI 53.5–110.4)82.0 ± 17.4 (CI 47.9–116.1)0.960
 3 61.0 ± 10.8 (CI 39.8–82.1)60.1 ± 13.5 (CI 33.6–86.6)0.453
 4 68.4 ± 12.6 (CI 43.7–93.0)68.0 ± 13.6 (CI 41.3–94.7)0.723
 5 90.7 ± 22.9 (CI 45.8–135.5)93.2 ± 27.2 (CI 39.9–146.5)0.147
 6 118.4 ± 31.8 (CI 56.0–180.7)118.1 ± 32.3 (CI 54.8–181.4)0.822
 7 118.4 ± 31.9 (CI 55.8–180.9)116.8 ± 29.8 (CI 58.4–175.1)0.269
 8 64.2 ± 14.1 (CI 36.5–91.8)64.9 ± 17.3 (CI 31.0–98.8)0.503
 9 52.8 ± 14.0 (CI 25.3–80.2)53.5 ± 14.2 (CI 25.7–81.3)0.382
 1071.8 ± 14.0 (CI 44.3–99.2)71.7 ± 13.6 (CI 45.0–98.4)0.975
 11 115.0 ± 29.5 (CI 57.1–172.8)113.6 ± 28.8 (CI 57.2–170.0)0.188
 12 114.8 ± 34.7 (CI 46.7–182.8)110.6 ± 39.1 (CI 34.0–187.2)0.175
Signal strength8.0 ± 0.7 (CI 6.6–9.3)8.1 ± 0.7 (CI 6.7–9.5)0.277

SD-OCT = spectral domain optical coherence tomography; AE = aligned exam; MAE = misaligned exam; AT = average thickness; CI = confidence interval. Student’s -test.