Research Article

Effects of Misalignments in the Retinal Nerve Fiber Layer Thickness Measurements with Spectral Domain Optical Coherence Tomography

Table 5

Comparison of mean ± standard deviation (95% confidence interval) SD-OCT parameters between eyes with and without misalignments within the measurement ring (sectors 2 and 3).

SD-OCTAE ()MAE () value

AT (µm)88.5 ± 14.5 (CI 60.1–116.9)88.6 ± 15.1 (CI 59.0–118.2)0.890
Quadrant, (µm)
 Superior 108.3 ± 21.0 (CI 67.1–149.5)107.8 ± 23.9 (CI 61.0–154.6)0.735
 Nasal 71.7 ± 11.0 (CI 50.1–93.3)72.4 ± 11.4 (CI 50.1–94.7)0.383
 Inferior110.4 ± 26.2 (CI 58.5–162.3)110.8 ± 26.6 (CI 58.7–162.9)0.831
 Temporal 63.3 ± 12.2 (CI 39.4–87.2)63.5 ± 13.6 (CI 38.6–92.0)0.844
Clock-hour, (µm)
 1 96.1 ± 23.3 (CI 50.4–141.8)97.2 ± 27.8 (CI 42.7–151.7)0.601
 2 83.8 ± 16.5 (CI 51.0–115.6)84.6 ± 17.2 (CI 50.9–118.3)0.583
 3 62.7 ± 11.2 (CI 40.7–84.7)62.9 ± 11.1 (CI 41.1–84.7)0.821
 4 68.5 ± 11.9 (CI 45.2–91.8)69.7 ± 13.7 (CI 42.8–96.6)0.301
 5 92.3 ± 23.1 (CI 47.0–137.6)95.8 ± 29.3 (CI 38.4–153.2)0.122
 6 122.2 ± 34.3 (CI 55.0–189.4)121.7 ± 35.4 (CI 52.3–191.1)0.813
 7 116.9 ± 32.3 (CI 53.6–180.2)114.8 ± 29.2 (CI 57.6–172.0)0.262
 8 64.1 ± 14.9 (CI 34.9–93.3)64.6 ± 19.2 (CI 27.0–102.2)0.716
 9 53.3 ± 16.1 (CI 21.7–84.9)53.5 ± 16.3 (CI 21.6–85.4)0.868
 1072.6 ± 14.8 (CI 43.6–101.6)72.2 ± 14.3 (CI 44.2–100.2)0.656
 11 116.5 ± 29.0 (CI 59.7–173.3)114.7 ± 28.9 (CI 58.1–171.3)0.185
 12 112.3 ± 27.1 (CI 59.2–165.4)111.3 ± 32.3 (CI 48.0–174.6)0.711
Signal strength8.0 ± 0.7 (CI 6.6–9.4)8.1 ± 0.8 (CI 6.5–9.7)0.378

SD-OCT = spectral domain optical coherence tomography; AE = aligned exam; MAE = misaligned exam; AT = average thickness; CI = confidence interval. Student’s -test.