Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry

1Instituto Balear de Oftalmología, 07011 Palma de Mallorca, Spain
2Departamento de Óptica II (Optometría y Visión), Facultad de Óptica y Optometría, Universidad Complutense de Madrid, 28037 Madrid, Spain
3Departamento de Bioquímica y Biología Molecular IV, Facultad de Óptica y Optometría, Universidad Complutense de Madrid, 28037 Madrid, Spain

Received 19 March 2015; Revised 28 May 2015; Accepted 28 May 2015

Academic Editor: Antonio Queiros

Copyright © 2015 Isabel Cacho et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Purpose. To compare the intraocular pressure (IOP) before and after Laser In Situ Keratomileusis (LASIK), measured by Diaton, Perkins, and noncontact air pulse tonometers. Methods. Fifty-seven patients with a mean age of 34.88 were scheduled for myopia LASIK treatment. Spherical equivalent refraction (SER), corneal curvature (K), and central corneal thickness (CCT) and superior corneal thickness (SCT) were obtained before and after LASIK surgery. IOP values before and after surgery were measured using Diaton, Perkins, and noncontact air pulse tonometers. Results. The IOP values before and after LASIK surgery using Perkins tonometer and air tonometers were statistically significant (). However, no significant differences were found () for IOP values measured with Diaton tonometer. CCT decreases significantly after surgery () but no statistical differences were found in SCT (). Correlations between pre- and postsurgery were found for all tonometers used, with and for the air pulse tonometer, and for Perkins, and and for Diaton. Conclusion. Transpalpebral tonometry may be useful for measuring postsurgery IOP after myopic LASIK ablation because this technique is not influenced by the treatment.